13

STARTING

Prior to starting the device, the system and process-specific safety measures must be observed.
For the media to be supplied, the relevant safety requirements and measures must be taken into
account.
Warning
Note
The following step-by-step procedure is recommended:
Check the set temperature on the built-in thermostat or on the external controller.
Warning
Switch on the power.
Note
Caution
www.mc-techgroup.com
Prior to starting the device, it must be verified that the system voltage
corresponds with the voltage shown on the rating plate!
If the set temperature on the capillary controller should be reduced in one
step by more than 30°C, the excess temperature circuit activates the
thermostat (press reset button to switch on again).
Caution: Do not touch the probe surface when in use.
The high surface temperatures that prevail are liable to cause burns.
Protective gloves must be worn and the probe must be protected to
prevent unauthorised access!
The total heating time is about 40 minutes. The probe is ready to use after
exceeding the lower alarm threshold value (30°C below the set value).
In the event of low temperature (failure of the probe heating), the sample
gas supply must be interrupted by suitable measures!
SP2500-H | 1.00.02
21