Performance Tests
40
Agilent 81630B/4B/6B/7B Power Sensor Modules and Agilent 81635A Dual Power Sensor Module, Second Edition
62 Decrease the attenuation of Atty2 by 10 dB in order to be on the
upper limit of the -50 dBm range.

-50 dBm Range

63 Disable Atty1.
64 Zero both Power Meters. On the 8163A/B with two installed power
meters, press [Menu], move to <Zero all>, and press Enter.
65 Enable Atty1
66 Switch one range up to the -40 dBm range and note the power
readings (#18).
67 Switch one range down (-50 dBm) and note the power readings
(#19)
68 Increase the attenuation of Atty1 by 10 dB and note the results in
the test record (#20).

Calculation

69 Calculate the non-linearity using the formulas given in the test
records.
Performance Test Procedures