Performance Tests
Performance Test for the Agilent 81637B
Page 3 of 3
Model Agilent 81637B Sensor Module
Test
No.
Test Description
I
Accuracy Test
measured at _______ nm
(1310nm)
measured at _______ nm
(1550nm)
II
Linearity Test
Range
P
DUT
+10
+9
+10
+3
0
+3
0
-7
-10
-7
-10
-17
-20*
-17*
-20
-27
-30
-27
-30
-37
-40
-37
-40
-47
-50
-47
-50
-57
III
Noise Test
IV
Return Loss Test
(V)
Relative Uncertainty due to Polarization
Maximum difference in interval
1550 nm - 1560 nm / 0.1 nm step; ∆P
(VI)
Relative Uncertainty due to Interference
Maximum difference in interval
1500 nm - 1560 nm / 2 pm step; ∆P
66
Agilent 81630B/4B/6B/7B Power Sensor Modules and Agilent 81635A Dual Power Sensor Module, Second Edition
Output Power
Output Power
For Calculations you may want to use the appropriate sheet
[dBm]
P
[dBm]
DUT
________
________
________
________
________
________
Reference
________
________
________
________
________
________
________
Report No.________
Min. Spec.
Result
[µW]
9.64 µW
_______
9.64 µW
_______
Loss [%]
_______
_______
_______
_______
_______
_______
0.0
_______
_______
_______
_______
_______
_______
_______
[pW]
_______
[dB]
55 dB <
_______
[dB]
_______
[dB]
_______
Test Record Sheets
Date ________________
Measurement Un-
Max. Spec.
certainty
10.36 µW
10.36 µW
<± 0.92 %
<± 0.46 %
<± 0.46 %
<± 0.46 %
<± 0.46 %
<± 0.46 %
<± 0.46 %
<± 0.46 %
<± 0.46%
<± 0.47 %
<± 0.47 %
<± 0.56 %
<± 0.56 %
<± 1.46 %
< 20 pW
< 0.01 dB
< 0.01 dB

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

81634b81636b81637b81635a