3.5 Test Results

When the test is complete, the device display will toggle between the sural nerve conduction amplitude
in microvolts (µV) and the conduction velocity in meters per second (m/s). In cases where the patient
has low amplitude (between 2-4 µV) and the CV cannot reliably be reported, only the amplitude will
display on the device. Amplitude result is indicated by a decimal point at the upper right corner of the
screen.
Recalling Results on Device
If the device powers down after a test, the results of that test can be recalled. To recall the results press
and hold the button for 3 seconds and you will see the results appear on the display. Results will display
for 10 seconds and then the device will power off. This mode can only be accessed when the device is
powered off. If you have powered on the device and need to recall the last result, wait 3 minutes for
the device to power down and then follow the instructions above for recalling results.
©2021 NeuroMetrix, Inc. All Rights Reserved. PN2203282 Rev N