As an alternative to shaft stubs, you can also use an additional adaptation
flange that is mounted on the sensor side and is included in the measurement
(see Fig. 4.6).
TB1A
Fig. 4.6: Additional adaptation flange for transfer transducers
To make it easy to install the gaging flange with the adaptation flange, we
recommend the following flange design:
• Provide clearance bores on the reverse of the flange to correspond to the
bore pattern of the TB1A. The relative position of the clearance bores
should correspond to the relative position of the tapped holes of the TB1A.
• The dimensions of flange clearance Z must be such that the connection
screws can be put through and tightened with a torque wrench.
A0162 7.0 en/de/fr
L
Adaptation flange
Z
D
On
L >
2
17
D
On