Using The Autoadapt Function; Switching Off Apnea Ventilation; Activating Autoadapt; Canceling Autoadapt - Dräger Carina Instructions For Use Manual

Subacute care ventilator
Hide thumbs Also See for Carina:
Table of Contents

Advertisement

Switching off apnea ventilation

1 Set menu 3.
2 Press the Apn.Vent. (B) key. Select Off by
turning the rotary knob and push to confirm.
The device ventilates with the previous ventilation
parameters.

Using the AutoAdapt function

The AutoAdapt function is used to adjust mask-
ventilated patients slowly to the therapeutic
pressure level.

Activating AutoAdapt

Prerequisites:
– SPN-CPAP/PS ventilation mode is activated.
– The Vent.set. menu is opened.
B C D E
1 Press the (A) key three times. Menu 4 is set.
2 Press the key for Aut.Adapt (E). Select On by
turning the rotary knob and push to confirm.
The AutoAdapt function is activated. The
parameters of the AutoAdapt function are
displayed:
– Aut.Adapt (B)
Δ
PSadpt (C)
Instructions for use Carina SW 3.2n
– PEEPadp (D)
– Tadpt (E)
3 Press the key of the corresponding parameter.
The ventilation parameters can be modified at any
time. AutoAdapt executes the initial adaptation with
reference to the modified ventilation parameters.

Canceling AutoAdapt

1 Set menu 4.
A
2 Press the key for Aut.Adapt (E). Select Off by
F
The device ventilates with the previous ventilation
parameters.
Select the value by turning the rotary knob (F)
and push to confirm.
turning the rotary knob and push to confirm.
Operation
67