Attaching Accessories To The Trolley - Dräger Carina Instructions For Use Manual

Subacute care ventilator
Hide thumbs Also See for Carina:
Table of Contents

Advertisement

3 Hold Carina (C) with both hands and position
the holes in the bottom of the device (D) on the
locating pins (E).
4 Move Carina parallel to the base plate (F) of the
device holder until it is resting safely on the
base plate.
5 Lock both locking levers: red areas are no
longer visible. Touch the locking levers on the
handle only.
6 Check that Carina is firmly seated.
Observe the maximum load, see "Trolley"
on page 126.

Attaching accessories to the trolley

Attaching the universal holder with standard
rail
1 Position the right side of the holder on the right
side of the column. Make sure that the catch (A)
of the universal holder is completely behind the
alignment aid.
C
B
2 Position the left side of the holder on the left
side of the column. Make sure that the catch (B)
of the universal holder is completely behind the
alignment aid.
Instructions for use Carina SW 3.2n
3 Use the template (C) on the top of the universal
holder to ensure that it is horizontal.
D
4 Align the universal holder so it is horizontal.
5 Tighten the adjusting screw (D).
CAUTION
Risk of damange to the device and of patient
injury
If the universal holder is not tightly fastened to the
column, it may come loose.
Check that the universal holder is tightly fastened
by pulling on it.
Height adjustment
Loosen the adjusting screw and adjust the
A
height of the universal holder. Align the
universal holder horizontally and tighten the
adjusting screw.
Placing accessories on the standard rail
Place accessories on the standard rail. Observe
the maximum load. See "Trolley" on page 126.
Assembly and preparation
37

Advertisement

Table of Contents
loading

Table of Contents