Advertisement

Assembly and preparation

Breathing circuit with leakage valve

The leakage valve can be used for both non-
invasive ventilation (NIV) and invasive ventilation
(Tube).
Setting the switch to leakage valve
B
1 Open the sliding cover (A) on the bottom of the
device.
2 Move the switch (B) to leakage valve
with a coin until the switch locks.
3 Close the sliding cover.
Connecting the breathing circuit
C
D
4 Fit the bacterial filter (C) to the inspiratory port
(D).
5 Connect the breathing hose (E) to the bacterial
filter.
6 Fit the leakage valve (F) to the port on the
breathing hose. Observe the direction.
When using the disposable breathing circuit
VentStar Carina LeakV (MP00312) the leakage
valve is already fixed to the breathing hose.
LeakV
The status line of the screen shows LeakV.
42
A
E
NIV
SPN-CPAP/PS

Breathing circuit with expiratory valve

NOTE
The breathing circuit with expiratory valve must
be used only for invasive ventilation (Tube).
Setting the switch to expiratory valve
1 Open the sliding cover (A) on the bottom of the
device.
2 Move the switch (B) to expiratory valve
with a coin until the switch locks.
3 Close the sliding cover.
Connecting the breathing circuit
G
C
F
H
D
4 Fit the bacterial filter (C) to the inspiratory port
(D).
5 Connect the breathing hose (E) to the bacterial
filter.
6 Fix the expiratory valve (F) to the port on the
breathing hose (E). Observe the direction.
When using the disposable breathing circuit
VentStar Carina ExpV (MP00313) the
expiratory valve is already fixed to the breathing
hose.
Instructions for use Carina SW 3.2n
A
B
E
F