UGO BASILE PAM Instruction Manual page 18

Pressure application measurement
Table of Contents

Advertisement

UGO BASILE - I
Press Measure. A screen will appear similar to:
PAM Transducer
Large: 0->1500gf
Calibrated: 091009
PAM gain matched
After a few seconds, the instrument completes an Autozero operation and then the display
shows.
IMPORTANT NOTE: do not apply force while the instrument is autozeroing.
Measure Peak Force
--------------------
Next
View
option config
Note that, in this case, all three soft buttons are in use.
Press Select this. The display is similar to:
23/07/09
14:23:35
Test 1
Exit
Check that no force is being applied to the connected transducer and then press Test.
PAM checks the transducer zero and then displays:
Measure Peak Force
Apply the force NOW
1
Apply gentle
force to the transducer. The display changes to typically.
For as long as the force is applied steadily, the Force value will indicate the applied force in
gf and the Time remaining will indicate the number of seconds left to complete the test.
Release the transducer after a few seconds and the screen changes to something like:
1
Do not apply more than 500 or 1500 gf (mouse or rat transducers respectively). Excessive force can
permanently damage the transducer.
TALY
Select
this
Test
Page 8
38500 Instruction Manual (Rev. 1)