Specifications
Time measurement functions
Waveform Math
Arithmetic
FFT
Triggering system
Trigger types
Trigger source selection
Horizontal trigger position
Aux In input impedance
Aux In maximum input voltage
Line trigger characteristics
6
Condition
Single shot, sample mode, full bandwidth
selected
> 16 averages, full bandwidth selected
The Sample Interval is the time between the samples in the waveform record.
Frequency, Period , Rise, Fall, Pwidth, Nwidth, Pduty, Nduty, DelayRR, DelayRF, DelayFR,
DelayFF, Burst width, Phase.
Add, Subtract, and Multiply waveforms
Spectral magnitude. Set FFT Vertical Scale to Linear RMS or dBV RMS, and FFT Window to
Rectangular, Hamming, Hanning, or Blackman-Harris.
Edge
Pulse Width
Runt
Analog channels, Aux In (External), AC Line.
The trigger position is set by the Horizontal Position knob.
1 MΩ ±2 %, 13 pF ±2 pF
300 V
, Installation Category II; derate at 20 dB/decade above 100 kHz to 13 V peak AC at
RMS
3 MHz and above
Based on sinusoidal or DC input signal. The maximum viewable signal while DC coupled is ±50 V
offset ±5 V/div at 4 divisions, or 70 V. AC coupling allows measuring signals on a DC level up to
300 V. For non sinusoidal waveforms, peak value must be less than 450 V. Excursions above
300 V should be less than 100 ms duration and the duty factor is limited to < 44%. RMS signal level
must be limited to 300 V. If these values are exceeded, damage to the instrument may result.
Line Trigger mode provides a source to synchronize the trigger with the AC line input.
Input Amplitude requirements: 85 V
Input Frequency requirements: 45 Hz - 440 Hz.
Time measurement accuracy
±(1 Sample Internal + 25 X 10
0.6 ns)
±(1 Sample Internal + 25 X 10
0.4 ns)
Positive or negative slope on any channel.
Coupling includes DC, HF reject, LF reject, and
noise reject.
Trigger on width of positive or negative pulses
that are >, <, =, or ≠ a specified period of time.
Trigger on a pulse that crosses one threshold
but fails to cross a second threshold before
crossing the first again.
- 265 V
.
AC
AC
TBS1000C Series Specification and Performance Verification
Time Base System (cont.)
-6
X |reading| +
-6
X |reading| +