Swan Analytical Instruments AMU Inducon Operator's Manual page 54

Version 6.20 and higher
Table of Contents

Advertisement

AMU Inducon
Program List and Explanations
5.3.2.342.2
5.3.2.342.8
5.3.2.44
5.3.2.54
5.3.2.6
5.3.2.7
5.3.2.1
5.3.4
5.3.4.1
5.3.4.2
5.3.4.3
52
Monday: Possible settings, on or off
to
Sunday: Possible settings, on or off
Run Time: see Interval
Delay: see Interval
Signal Outputs: see Interval
Output/Control: see Interval
Function = Fieldbus:
The relay will be switched via the Profibus input. No further param-
eters are needed.
Input: The functions of the relays and signal outputs can be de-
fined depending on the position of the input contact, i.e. no function,
closed or open.
Active: Define when the input should be active:
The measurement is interrupted during the time the input is active.
Input is never active.
No:
Input is active if the input relay is closed
When closed
Input is active if the input relay is open
When open:
Signal Outputs: Select the operation mode of the signal outputs
when the relay is active:
Signal outputs continue to issue the measured value.
Cont.:
Hold:
Signal outputs issue the last valid measured value.
Measurement is interrupted. Errors, except fatal
errors, are not issued.
Set to 0 or 4 mA respectively. Errors, except fatal
Off:
errors, are not issued.
Output/Control: (relay or signal output):
Controller continues normally.
Cont.:
Controller continues on the last valid value.
Hold:
Controller is switched off.
Off: