4.5.2 Secondary Emergency Release

A RED Emergency Grip
It is recommended that the primary emergency release be
used. The secondary emergency release is only a back-up to
the primary emergency release.
In cases where the primary release is either not functioning or
unreachable, a secondary emergency release may be used.
Pull up on the EMERGENCY grip A and push down on the lift
1.
arms B at the same time.
1164940-D
B Lift Arms

4.6 Performing an Emergency Stop

Press the RED button A on the control box B in to stop the
1.
boom and patient from raising or lowering.
2.
To reset, rotate the emergency button clockwise.

4.7 Charging the Battery

Invacare recommends the battery be recharged daily to
prolong battery life.
There are two different methods to charge the battery. One
method uses a power cord that attaches to the control box,
the other requires the battery to be mounted to the battery
charger. Follow the appropriate procedure to charge the
battery for your patient lift.