ABB MAG-SM Instruction Bulletin page 29

Field mount housing and 19“-design
Table of Contents

Advertisement

Submenu/Parameter
Submenu
Function test
ENTER
Electromagnetic Flowmeter Converter
Entry Type
Slave-Addr.
126
Function
Param.-Profib. DP
Communication
ASCII-SM1 mode
Communication
µDCI-Binary
Instr. address
004
Baudrate
2400 Baud
Printer type
Standard
Print time
Function test
I
out
Function test
VNVRAM, 22C12
Comments
The Slave-Address must be entered as a three digit number,
address range 000, 001 to 126. If a bus address is entered as a
one or two digit number incorrect interpretation of the bus address
by the converter will result.
This function is used to access the parameters from the Profibus
DP-Module. See also Section 13 „Functions in the Data Link
Menu" in the Profibus Data Link Description, Part Number
D184B093U03.
ASCII-SM1000-Mode
In this protocol the contents of the Error Register can be
transmitted bitwise.
µDCI-Binary-Protocol
Compatible with the µDCI-Family (process automation)
Instrument Address: 0–99 (not available when Printer
protocols are selected). If multiple instruments are connected to a
single bus (RS 485), each instrument connected to the bus must
have a unique address.
Baudrate: Selection range 110–28800 Baud.
This parameter is only displayed when a printer protocol has been
selected.
1)
Either a standard printer type
ABB 55DE1000 can be selected.
This parameter is only displayed when a printer protocol and the
ABB 55DE1000 Printer have both been selected.
Entries for the Protocol Printer:
Year, Month, Day, Hour, Minute
Exit the submenu
C/CE
Function test current output, enter data mA.
Function test Fout
See Section. 5.4 on Page 27
Function test internal subassemblies, automatic tests of NVRAM,
EPROM 27C512, EEPROM 93C46,
ext. EEPROM 93C46. Additional test functions:
Alarm contact, F/R-contact, terminals P1/P2, terminals P3/P4,
input A1, input A2, data link, Profibus, F
Ext. zero return and totalizer reset.
1) Minimum requirements for a Standard Printer
40 characters/line, 1 kByte printer buffer,
compatible ASCII-character set.
Handshaking (e.g. XON/XOFF) is not utilized.
A print output of the report is initiated over the ext. totalizer
reset contact input, terminals G2/U2-31.
or the Report Printer
, display,
out
23