ABB MAG-SM Instruction Bulletin page 26

Field mount housing and 19“-design
Table of Contents

Advertisement

Electromagnetic Flowmeter Converter
Submenu/Parameter
Pulse
1.000
/m3
Pulse width
30.000 ms
Low flow cutoff
1.000 %
Damping
10.0000 s
Filter
on
Density
2.54300
g/cm3
System zero
3.5 Hz
20
Entry Type
numeric
numeric
numeric
numeric
from table
numeric
from table/numeric
Adjust?
yes → ENTER
ENTER
Adjust
Manual
Adjust
Automatic
Comments
Pulse factor for int. and ext. flow totalization,
Range 0.001–1000 pulses per selected flow unit,
max. count frequency 10 kHz.
Note: The max. count frequency and the pulse factor are checked
by the software, if out of range Error 40
"Frequency >10 kHz" is displayed.
For external pulse output. Range: 0.0032 ms – 2000 ms in
multiples of 0.0032 ms. The max. allow. pulse width is checked
by the software and corrected if required.
Range 0–10 % of the flow range end value setting, applies to the
displayed values and all outputs.
The switching limits for the zero return incorporate a
hysteresis of 0.5 %.
Range 0.100–99.99 s. Response time for 5
flowrate change.
Noise Signal
Without Filter
On/off. Standard is off. If the output signal is noisy, turn the filter
on and select a damping time > 2.4 sec. For applications with
piston or diaphragm pumps (pulsating flow) a special algorithm is
required. Faster flowrate measurements are achieved at this
higher excitation frequency and utilizing difference totalizers so
that the reverse flow values can be subtracted, i.e. the totalizer
runs in reverse. The filter should be set to "On" for pulsating flows
applications (for a pulsation free output).
3
Range 0.01–5 g/cm
. Mass flowrate display and totalization
in g, kg, t, uton or pounds.
Exit the submenu
C/CE
Zero value check (required when retrofitting older flowmeter
primaries).
Manual entry, e.g. when a converter is exchanged.
Valve must be closed, flowrate must be at 0. The automatic
adjustment is initiated with ENTER. The zero value limits are
±1500 Hz. If the value exceeds these limits, the zero adjustment
value is not accepted.