Performing manual disinfection
WARNING
Liquids entering the device can disturb the
function of the device or damage the device
and endanger the patient! Wipe device
surfaces and cables with disinfectant using
only moist cloths and make sure that no
liquids enter the device.
1 Disinfect surfaces.
2 Remove any disinfectant residues after the
contact time.

Maintenance

Maintenance work to be carried out only by
specially trained service personnel
no later than every 6000 operating hours (see
operating hours counter):
 Change filters in prefilter, main filter and intake
filter. Dispose of used filters with domestic
waste.
Carry out maintenance more frequently if
necessary.
If the device has not been in operation for
6000 hours:
 Carry out inspection and servicing after not
more than one year.
30

Cleaning the filter mat

A
1 Regularly check the filter mat for soiling.
If the filter mat is soiled:
2 Remove the filter mat (A), knock out any dirt,
insert the filter mat again.
Instructions for Use Air Compressor