Fluke 1630-2 Calibration Manual page 16

Earth ground clamp
Hide thumbs Also See for 1630-2:
Table of Contents

Advertisement

1630-2/1630-2 FC
Calibration Manual
3.
At the next calibration step, if the screen shows Ω, apply a resistance loop to the jaw with the precision
decade resistance box an a 1-loop short copper wire. See Figure 2. The applied resistance value should be
the same on the calibrator as the value on the screen below.
12
Figure 2. Calibration with 1-Loop Short Copper Wire
IET 1433-20
H
L
G