9060 High Voltage Controller - Operation
N O T E
Some applicators, ES (Evolver SE, Esta-
Quick) and Ab (Aerobell 168), have adjust-
able overload values. Please refer to the
"Parameter Adjustment Menu" part of the
"Operations" section for information on ad-
justing the overload value.
KV TEST JUMPER
To assist in testing and troubleshooting, a
jumper (J8) has been added to the main PC
board. By covering (shorting) both terminals of
this jumper, the high voltage of the spray
applicator can be activated. Thus, for testing
and troubleshooting, high voltage output can be
obtained without the need to trigger air through
the spray applicator. After testing, the jumper
must be repositioned so that it covers only
one terminal (open) or the high voltage will
stay on all the time. See Figure 25 for the
location of test jumper J8. Ensure the jumper is
installed on ONLY ONE post during power on to
prevent a boot fault "bF".
Figure 25: KV Test Jumper Location
W A R N I N G
!
If jumper J8 is left covering (shorting)
both terminals, the High Voltage will re-
main on or cause a boot fault (bF) when
the unit is powered on.
N O T E
Use Ransburg Calibrated Equipment
ONLY for testing and troubleshooting. Re-
fer to the "Accessories" section for this
manual for part numbers for testing equip-
ment.
33
REMOTE I/O MONITORING
DIAGNOSTIC MODE
The remote I/O monitoring diagnostic mode has
three (3) sub modes.
modes provide the ability to monitor the state of
the analog and digital signals. The third sub
mode displays the gun type value and firmware
version.
To enter the remote I/O monitoring
and diagnostic mode, ensure that the high
voltage is not activated and press all 3 preset
buttons (1, 2, and 3) at the same time. This will
put the unit into the Analog Signal Monitoring
sub mode, with the preset 1 LED indicator lit.
Each sub mode is entered by pressing one of
the three preset buttons, and the corresponding
LED indicator will light up. The following table
lists which sub mode corresponds to which
preset. To exit the diagnostic mode, from any
sub mode, press the Reset Button.
Remote I/O Diagnostic Modes
Preset Button
Preset 1
Analog Signal Monitoring
Preset 2
Preset 3
Gun Type / Firmware Version
N O T E
Remote I/O Monitoring Diagnostic Mode
can ONLY be entered when the High Volt-
age is OFF and is ONLY available for the
ES (Esta-Quick, Evolver SE) and Ab
(Aerobell 168) applicators.
While the unit is in remote I/O monitoring
and diagnostic mode, the High Voltage will
NOT be able to be activated via the trigger
signal.
N O T E
Each time a signal monitoring sub mode
is entered, the signal index is reset to zero
(0) moving it back to the start of the signal
list for the selected mode.
The first two (2) sub
Sub Mode
Digital Signal Monitoring
CP-13-05.3