L011-57 (Rev D1, 2018-11-09) Contains directions for the following products: MT-LPP, MT-LPP-APM, 66-EL-LP www.fh-co.com FHC, Inc. FHC Europe FHC Latin America 1201 Main Street (TERMOBIT PROD srl) Calle 6 Sur Cra 43 A-200 Bowdoin, ME 04287 USA 42A Barbu Vacarescu Str, 3rd Fl...
Table of Contents: Indications for use, Intended use, Applicable Standards Symbol Key Warnings and Cautions System Maintenance and Repair Handling Considerations Maintenance Repair System Disposal System Specifications - Performance & Accuracy Microelectrode Recording Stimulation Impedance Check Analog Inputs Inventory Ordering Information System Set-up and Pre-Use Checkout L011-57 Rev.
WARNING: The Guideline 4000 LP+ issues no alarms should undesirable or dangerous medical conditions arise. WARNING: The Guideline 4000 LP+ must be operated by a person who has been trained by an FHC authorized representative and has read and understood the directions for use (DFU).
WARNING: Do not allow any liquid to contact any Guideline 4000 LP+ component. Maintenance • The Guideline 4000 LP+ must be inspected and recalibrated each year by an FHC authorized representative. Contact FHC for more information on Service Agreements and Preventive Maintenance Programs. •...
Stimulation Mode: Constant Current, Compliance Voltage: ±50V Microstimulation: Stimulus Parameters Macrostimulation: Stimulate from microelectrode to common From macro-contact to common or between • Monophasic or biphasic, cathode or macro contacts • Range: 0 to 100μA, Display Resolution: 1 μA, anode leading or user defined Accuracy: ±5%, see performance graphs •...
CAUTION: If using a shipping case with a lock, do not lose the key for the lock. WARNING: Packing foam can not be sterilized. If contamination of the foam is suspected, contact FHC for disposal instructions. CAUTION: Sharp edges, use caution.
Page 8
WARNING: Do not enable the wireless capabilities of the notebook computer in the hospital environment unless you are authorized to do so. WARNING: Do not install third party software onto the Guideline 4000 LP+ notebook computer unless authorized to do so by an FHC factory authorized representative.
Page 9
16. Verify the impedance check and stimulation functionality of each available channel (by repeating step 3B) before every surgery. WARNING: If an error occurs during the pre-use checkout procedure (step 3), discontinue use of the system. Contact an FHC factory authorized representative for further instructions.
Page 10
If using the microTargeting Controller and motor or encoder, connect the microTargeting motor or encoder assembly as shown and perform the following steps: Zero the microTargeting Controller by pressing the small black button on the front panel of the Controller. The depth readout of the microTargeting Controller should appear on the computer screen.
Page 11
WARNING: Operation of the Guideline 4000 LP+TM in proximity of short wave or microwave therapy equipment may produce instability in the electrical stimulator output. WARNING: Do not arrange electrodes and equipment in such a way that stimulation current may pass through the heart.