System Maintenance and Repair:

Handling Considerations

The Guideline 4000 LP+ is intended to be used in a standard operating room environment.
Store the Guideline 4000 LP+ at room temperatures between 0°C (32°F) and 40°C (104°F).
Neither the Guideline 4000 LP+ nor its components are designed for sterilization.
Do not drop the Guideline 4000 LP+ or its components.
Only transport the Guideline 4000 LP+ inside of its shipping container, with included foam insulation.
WARNING: Do not expose the Guideline 4000 LP+ to humidity of greater than 95%.
WARNING: Do not store the Guideline 4000 LP+ near or expose it to strong magnetic fields.
WARNING: Do not allow any liquid to contact any Guideline 4000 LP+ component.

Maintenance

The Guideline 4000 LP+ must be inspected and recalibrated each year by an FHC authorized representative. Contact FHC for
more information on Service Agreements and Preventive Maintenance Programs.
The Guideline 4000 LP+ and its components may be cleaned by wiping a dry rag lightly over all surfaces. If necessary, a slightly
damp cloth moistened with a standard hospital disinfectant may be used.
CAUTION: Only qualified FHC representatives may calibrate the Guideline 4000 LP+.

Repair

All FHC products are unconditionally guaranteed against defects in workmanship for one year from the date of shipment,
provided they have been exposed to normal and proper use.
Should service or repair be required, please contact FHC at 1-800-326-2905 (US & Canada) or +1-207-666-8190 for return
instructions.
End-of-Life System Disposal
Return the Guideline 4000 LP+ to FHC for proper end-of-life disposal. Please contact an FHC factory authorized
representative to arrange a return.
Password Reminder:
System Specifications - Performance & Accuracy:

Microelectrode Recording

Analog signal conditioning
• Gain: 54 to 7201 user adjustable,
Amplitude accuracy: ±5%, Polarity: Inverting
Upper frequency cutoff: 20kHz, 40dB/decade
roll-off
• Lower frequency cutoff: MER - 200Hz, 20dB/
decade; LFP - <1Hz, 20dB/decade
L011-57
Digital signal processing
• Sampling: 16 bit resolution over ±1.4V range
• Sample Rate: MER: 10-48kHz ,LFP: 5-1000Hz
Digital Bandpass Filter: High Pass - 0 to 1000Hz,
Low Pass – 100 to 20kHz
Adaptive line noise filter: adaptation weight
adjustable from 0 to 10% per cycle
Adaptive Stimulus Artifact Suppressor: adaptation
weight adjustable from 0 to 10% per cycle
Timing accuracy: ±10ppm (0.001%), resolution:
1/48,000 s (~20.8µs)
Rev. D1
Waveform Analysis
• Display Mode: scroll/sweep rate
• Spike discrimination method: Dual line
• Firing Rate: displayed in Hz, one second
2018-11-09
adjustable from 5s/window to 0.25s/
window
Spikes Mode: duration 290μs/sweep to
2.585 ms/sweep in 12 steps
level or Window
refresh rate
Page 5 of 11