Chapter14
TROUBLESHOOTING

14.1 Powering off Automatically

1 Whether or not the battery capacity is nearly exhausted? Battery
overdischarge protection circuit is activated.
2 Whether or not AC supply voltage is too high? Overvoltage protection
circuit is activated.
3 Whether or not AC interference is too large or fixed knob in lead cable
socket is tightened? Overload protection circuit is activated.

14.2 AC Interference

1 Whether or not the device is grounded reliably?
2 Whether or not the electrodes or lead cables are connected correctly?
3 Whether or not the electrodes and skin are smeared enough electric gel?
4 Whether or not the metal bed is grounded reliably?
5 Whether or not the patient has touched the wall or metal parts of the bed?
6 Whether or not the patient has touched other people?
7 Whether or not there is high-power electric equipment working nearby?
Such as X-ray machine or ultrasonic device, etc.