ISA T 1000 Plus Introductory Manual
Hide thumbs Also See for T 1000 Plus:
Table of Contents

Advertisement

DATE: 27/11/2011
DOC.MIE92093
REV. 1.35
T 1000 PLUS
INTRODUCTORY GUIDE

Advertisement

Table of Contents
loading

Summary of Contents for ISA T 1000 Plus

  • Page 1 DATE: 27/11/2011 DOC.MIE92093 REV. 1.35 T 1000 PLUS INTRODUCTORY GUIDE...
  • Page 2 DOC. MIE92093 Rev. 1.34 Page 2 of 108 REVISIONS SUMMARY VISA PAGE DATE 20/10/2008 Issued Lodi 1.28 26 2/02/2010 Added the paragraph Lodi test results handling 1.28 17, 37 1/10/2010 Added TD1000 Lodi model and the 15 Hz 1.34 11 –13; 7/9/2011 Revised to issue 1.34;...
  • Page 3: Table Of Contents

    DOC. MIE92093 Rev. 1.34 Page 3 of 108 SHORT FOREWORD ..................5 INTRODUCTION ..................... 6 1 TEST SET EXPLANATION ..............9 1.1 C ONNECTION TO THE RELAY AND POWER ........... 9 1.2 T EST CONTROL ..................10 1.3 C URRENT GENERATION ................
  • Page 4 4.16.1 Introduction ................96 4.16.2 Calibration procedure ............. 96 4.16.3 T 1000 PLUS output calibration ........ 97 4.16.4 T 1000 PLUS external meas. calibration ....99 4.16.5 Phase angle measurement calibration ....100 APPENDIX 1 SPARE PARTS LIST ............. 105...
  • Page 5: Short Foreword

    So, either the manual is actually of help, or I ignore it. This is why I decided to split the T 1000 PLUS manual in three: specification, with all performance details; application manual, with instructions about how to use it one its operation is understood;...
  • Page 6: Introduction

    Page 6 of 108 I NTR O D U CT IO N The single phase relay test set mod. T 1000 PLUS is suited for the testing and adjustments of the following types of relays; the table lists also the paragraph that explains the test procedure.
  • Page 7 PC at a later time, along with settings. The ease of operation has been the first goal of T 1000 PLUS: this is why the LCD is graphic, and so large. With it, the dialogue in MENU mode is made easy.
  • Page 8 Additional features are: . Two meters, current and voltage, with independent inputs, allow measuring T 1000 PLUS outputs or any other source; . Two auxiliary contacts, that allow simulating the circuit breaker; . A set of resistors allows easing output adjustment.
  • Page 9: Test Set Explanation

    . Auxiliary supply voltages. If something is wrong, the operator is alerted by a message. At the end of it, default selections are active; T 1000 PLUS is in the OFF state. Perform the first selections, according to the type of relay to be tested: .
  • Page 10: Test Control

    1.2 T EST CONTROL The T 1000 PLUS front panel is explained in next paragraph. T 1000 PLUS generation is controlled by the two keys < (55) and > (56). Settings and menu selections are controlled by the multi-function knob with switch (22): see next paragraph for menu selections description.
  • Page 11 Test results can be saved according to Save selections. . From ON to OFF + TIME: main outputs are removed the timer starts according to selections; as STOP is sensed, T 1000 PLUS returns OFF, the TRIP LED (43) turns on and parameters at trip are displayed until ON or ON+TIME are selected.
  • Page 12 DOC. MIE92093 Rev. 1.34 Page 12 of 108 FAUL MAX TIME RELAY 0.1 s DELAY . Trip + pulse time: the timer measures the delay and the duration of the trip impulse. . Reclose test. It is possible to select via menu the test of a reclosing scheme.
  • Page 13 DOC. MIE92093 Rev. 1.34 Page 13 of 108 FAULT TRIP (STOP) RECLOSE (START) TIME MEAS. Figure 4: Measure of Delay and Reclose times The second operating mode refers to pole mounted CB’s. In this mode, there is only one signal coming from the device under test: the position of the CB.
  • Page 14: Current Generation

    DOC. MIE92093 Rev. 1.34 Page 14 of 108 . Timed: in all modes (ON; ON+TIME; OFF+TIME), fault outputs are generated for the programmed maximum time; after this, T 1000 PLUS returns OFF. Any trip after this time is not sensed. .
  • Page 15 DOC. MIE92093 Rev. 1.34 Page 15 of 108 1) MAXIMUM POWER 300 VA RANGE CURR. MAX. MAX. LOAD RECOV. A AC OUTPUT POWER BURDEN TIME TIME 0.33 STEADY 0,17 30 min 0.075 0.04 1000 0.016 STEADY 30 min 0.66 0.25 1000 0.15 STEADY...
  • Page 16 The procedure is the following. . At first, be sure that the main control knob (6) is turned (rotated) to the zero position (complete counter-clockwise). . Power-on T 1000 PLUS.
  • Page 17 DOC. MIE92093 Rev. 1.34 Page 17 of 108 . Select by the push-button (57) the measurement on the desired output sockets (13), according to the maximum current to be generated: the LED turns on; the AC voltage value is displayed. .
  • Page 18: Ac Voltage Generation From Main Output

    . Connect the relay to be tested to sockets (60). Check that the adjusted voltage does not drop as you connect the relay; else, this would mean that T 1000 PLUS is overloaded (or that you are connecting to a live wire). In this situation, remove the cause of error and connect again.
  • Page 19: Ac Voltage Generation From The Auxiliary Output

    . Connect the relay to be tested to sockets (61). Check that the adjusted voltage does not drop as you connect the relay; else, this would mean that T 1000 PLUS is overloaded (or that you are connecting to a live wire). In this situation, remove the cause of error and connect again.
  • Page 20 Check that the adjusted voltage does not change or the overload message pops up as you connect the relay; else, this would mean that T 1000 PLUS is overloaded (or that you are connecting to a live wire). In this situation, remove the cause of error and connect again (reset the alarm if it popped up).
  • Page 21 DOC. MIE92093 Rev. 1.34 Page 21 of 108 pre-fault voltage, that simulates the situation prior to fault, and the fault voltage. The pre-fault voltage adjustment is performed by the control knob, while knob (20) adjusts the fault value. Voltage output selection is automatic: pre-fault voltage with test stopped;...
  • Page 22 > Phase (phase) ESC This angle is referred to the fault voltage; so, for its adjustment is not necessary to have T 1000 PLUS ON. Last, it is possible to test frequency rate of change relays, by setting both the starting frequency, as before, and the frequency ROC range, with range from ±...
  • Page 23: Dc Voltage Generation From The Auxiliary Output

    (63), after pressing the push- button (69). Use this generator as follows. . Power-on T 1000 PLUS: the output voltage is zero. Press the push-button (69) to enable or disable the output; when enabled, light (63) is ON, and the DC voltage value is displayed on the screen.
  • Page 24: Auxiliary Contacts

    Rev. 1.34 Page 24 of 108 relay; else, this would mean that T 1000 PLUS is overloaded (or that you are connecting to a live wire). In this situation, if it is not a connection error, it is possible to reduce the voltage until the voltage does not drop: relays tolerate a wide range of DC supply voltages.
  • Page 25 . Timer start: internal; as test is started. External start allows synchronizing more T 1000 PLUS. . Timer stop: external. Internal stop means that the timer is stopped as T 1000 PLUS goes OFF. With this selection, the timer meters the elapsed time between START and STOP. Alternative selections: ..
  • Page 26: Finding Relay Thresholds

    DOC. MIE92093 Rev. 1.34 Page 26 of 108 the corresponding energy (if selected). The selection is performed as follows: TIMER START/STOP > STOP > External > Count > (number of counts) ESC If the count is selected on START, time measurement will be performed after the set number of counts has expired: this serves to pass the start-up of the energy meter.
  • Page 27: First Threshold Trip And Drop-Off

    DOC. MIE92093 Rev. 1.34 Page 27 of 108 1.10.2. First threshold trip and drop-off The first session is finding threshold I>, that is the limit between no trip and trip (with long delays). With any other type of single- threshold relay, the procedure is the same, provided that the corresponding parameter is changed.
  • Page 28: Second Threshold Trip And Drop-Off

    DOC. MIE92093 Rev. 1.34 Page 28 of 108 1.10.3. Second threshold trip and drop-off The second session is finding threshold I>>. The problem is that the test result criterion is no more to find the limit between no trip and trip; it is instead to find the limit between two different timings: what we have shown as t>, for currents less than I>>, and t>>...
  • Page 29 DOC. MIE92093 Rev. 1.34 Page 29 of 108 . Under-current relay trip delay: from ON to OFF+TIME; . Under-current relay drop-off delay: from OFF to ON+TIME. All these tests are performed after having pre-adjusted the desired test current (or set of parameters). FAULT Stop Trip delay...
  • Page 30: Est Results Handling

    DOC. MIE92093 Rev. 1.34 Page 30 of 108 . Now, press ON and pre-adjust the first set of fault parameters: as the relay trips, don’t save test result; go OFF. . Select ON+TIME: as the relay trips, test goes OFF; pressing the multi-function knob tripping values can be saved.
  • Page 31 DOC. MIE92093 Rev. 1.34 Page 31 of 108 The operation to input the header is the following.  After having entered, if you move the knob it will move between: arrow up; Plant name; arrow down, Return.  If you press the arrow up or down, you will scroll through: Plant name, Operator, Serial number, Model/manuf., Feeder.
  • Page 32: Basic Test Principles

    DOC. MIE92093 Rev. 1.34 Page 32 of 108 sure to save the correct data. Save is performed both in ON mode and in ON+TIME mode. With the selection Manual, it is possible to save data at any time, pressing MENU and then Yes, even if the relay did not trip. As already explained, a threshold is verified with two tests: with a value the relay trips, with a slightly different value the relay does not trip.
  • Page 33: Parameter Vs. Time Characteristic

    DOC. MIE92093 Rev. 1.34 Page 33 of 108 The easiest type of test is to check the settings. In this instance it is sufficient to generate the relevant parameters (voltage, current, angle) with values close to the settings. With two timing tests, little below and little above the settings, it is possible to test that they are correct, within a specified tolerance.
  • Page 34: Parameter Vs. Parameter Characteristic

    DOC. MIE92093 Rev. 1.34 Page 34 of 108 t>> I> I>> I-t over-current relay curve The curve means that: . If the test current is less than I> no trip occurs; . For current more than I> and less than I>> trip time is a function of the test current;...
  • Page 35 DOC. MIE92093 Rev. 1.34 Page 35 of 108 keeping the other one constant. However, we must be careful before executing the test. Vmin Imin I-V directional relay curve For instance, in order to find the point A it is apparent that the procedure is: .
  • Page 36 DOC. MIE92093 Rev. 1.34 Page 36 of 108 M I N The area to the right is the trip zone; to the left it is the no trip zone. The test strategy should be the following: . For points A, B, D: set the test voltage and increase the current until the relay trips;...
  • Page 37: Use Of The Test Set As A Multimeter

    DOC. MIE92093 Rev. 1.34 Page 37 of 108 The area of intervention is the one within the circle with the centre in the negative X axis. The first problem is to convert R and X into the corresponding V, I, angle; then, the current is set at IN, and only voltage and phase are modified.
  • Page 38 DOC. MIE92093 Rev. 1.34 Page 38 of 108 . The frequency is measured only on the voltage input: connect it and select the F measurement. . The measurement of the impedance or of the resistance of a component fed by an external source is performed as follows. ..
  • Page 39: Output

    DOC. MIE92093 Rev. 1.34 Page 39 of 108 N. PARAMETER , AC DERIVED FORMULA UNITS INPUTS FROM ACTIVE POWER, P Iext, Vext; P = I*V*cos φ (φ) REACTIVE POWER, Q Iext, Vext; φ I*V*sin(φ) APPARENT POWER, S Iext, Vext S= I*V POWER FACTOR, p.f.
  • Page 40 DOC. MIE92093 Rev. 1.34 Page 40 of 108 Hz: these values are higher than those of the standard T1000 PLUS. RANGE @ VMAX POWER @ VMAX POWER @ 50 Hz 15 Hz 15 Hz 33.3 Hz 33.3 Hz VOUT VOUT 20 @ 20 A;...
  • Page 41 DOC. MIE92093 Rev. 1.34 Page 41 of 108 Next, all you have to do is to connect FT1000 in series to the T1000 selected output: the FT1000 range shall be the same or greater than the one of T1000.
  • Page 42: Test Set And Pop-Up Menu

    DOC. MIE92093 Rev. 1.34 Page 42 of 108 2 TEST SET A N D P O P - UP M E N U 2.1 T HE FRONT PANEL...
  • Page 43 DOC. MIE92093 Rev. 1.34 Page 43 of 108 The following list includes the key components inside T 1000 PLUS; see the schematic on last page. 1) Main supply fuse, rated T10A, incorporated in the supply socket. 2) Power-on switch. 6) Main outputs adjustment. 8) Earth socket.
  • Page 44: Display And Control Lights

    DOC. MIE92093 Rev. 1.34 Page 44 of 108 52) OFF + TIME light: current is removed and time metered until STOP is detected. 53) Main AC voltage selection light. 54) Main DC voltage selection light. 55 and 56) START and STOP push-buttons. 57) Push-button for the selection of main output.
  • Page 45: The Pop - Up Menu

    PC provided with TDMS. The software allows saving test results, examining them and so on. The specification of TDMS is given in a separate document. When the PC is connected, settings can also be created and transferred into T 1000 PLUS using TDMS.
  • Page 46 DOC. MIE92093 Rev. 1.34 Page 46 of 108 The flux diagram of menu selections can be found in Appendix 1. LEVEL1 LEVEL 2 LEVEL LEV. 4 FUNCTION TEST Test Normal Measures the time CONTROL mode (default) delay from START (internal, external) to STOP (internal, external).
  • Page 47 Rev. 1.34 Page 47 of 108 External Generation starts upon reception of the START input: this allows synchronising T 1000 PLUS. When selected, the fault is injected as soon as the selected condition on START input is met. Timed Generation lasts for...
  • Page 48 DOC. MIE92093 Rev. 1.34 Page 48 of 108 TIMER Start INT (default) Timer starts when ON START/ or ON+TIME are STOP activated and outputs generated. NO-NC- After ON or ON+TIME, EDGE timer starts on the external input. External START input Normally Open, or Normally Closed, or Both (EDGE).
  • Page 49 DOC. MIE92093 Rev. 1.34 Page 49 of 108 CLEAN- Timer stops when the 24 V – 80 STOP input is detected. External STOP input without or with voltage. If with voltage, two voltage thresholds are available: 24 or 80 V. COUNT Timer enters the counting mode;...
  • Page 50 DOC. MIE92093 Rev. 1.34 Page 50 of 108 LEVEL LEV. LEV. LEVEL LEVEL 5 FUNCTION Range 65 (default) ; VAC/ 130 ; 260 V. control Mode Fault (default) The auxiliary AC voltage is adjusted by the dedicated knob, and is available after pressing the ON-OFF...
  • Page 51 DOC. MIE92093 Rev. 1.34 Page 51 of 108 Prefault Prefault Sets the pre- + Fault Amplitud fault auxiliary AC voltage amplitude. Entering this selection in OFF mode, the pre-fault voltage is generated and displayed, and adjusted by the multi- function knob. NOTE: the fault voltage is generated...
  • Page 52 DOC. MIE92093 Rev. 1.34 Page 52 of 108 Prefault Sets the duration duration of the pre-fault auxiliary voltage. When ON or ON+TIME are pressed, the pre-fault will be generated at the mains frequency for the selected duration; then the fault voltage is generated, at programmed...
  • Page 53 DOC. MIE92093 Rev. 1.34 Page 53 of 108 Adjust The frequency freq 500.00 of the auxiliary voltage may be programmed. Frequency changes at test start; output voltage does not change in amplitude. Adjust The frequency r.o.c.: 0.01.. ramps at the 9.99 programmed Hz/s...
  • Page 54 DOC. MIE92093 Rev. 1.34 Page 54 of 108 Adjust phase Vaux The fault - mains auxiliary voltage can be phase shifted with respect to the mains. The measured angle is displayed. Test must be ON; for a correct angle measurement, the auxiliary voltage must be more than...
  • Page 55 DOC. MIE92093 Rev. 1.34 Page 55 of 108 Adjust phase Vaux The fault – I main auxiliary voltage can be phase shifted with respect to the main current. The measured angle is displayed. Test must be ON; for a correct angle measurement, current and...
  • Page 56 DOC. MIE92093 Rev. 1.34 Page 56 of 108 Adjust phase Vaux The fault – V main auxiliary voltage can be phase shifted with respect to the main voltage. The measured angle is displayed. Test must be ON; for a correct angle measurement, both voltages...
  • Page 57 DOC. MIE92093 Rev. 1.34 Page 57 of 108 LEVEL1 LEVEL 2 LEVEL 3 LEVEL 4 FUNCTION METERS Internal Units of I Normal If selected, current values are displayed in I/IN If selected, displayed values are defined as I/IN, that can be defined.
  • Page 58 DOC. MIE92093 Rev. 1.34 Page 58 of 108 10A – 20 Selects the current input socket Waveform If selected, the current waveform is displayed Externa Enabled With (default) - selection AC the meter performs the true rms measureme nt; with selection measureme nt is...
  • Page 59 DOC. MIE92093 Rev. 1.34 Page 59 of 108 LEVEL1 LEVEL 2 LEVEL 3 FUNCTION METERS Other None (default) No extra (continue internal measurement displayed Active power P; W Reactive power Q; VAr Impedance module Z, Ohm Impedance argument φ, ° Active impedance R, Ohm component...
  • Page 60 DOC. MIE92093 Rev. 1.34 Page 60 of 108 Note: measurements marked AC apply only if both inputs are selected as alternate current. LEVEL1 LEVEL 2 LEVEL 3 LEVEL FUNCTION RESULTS Test header Groups results Show results Display result Delete Selected result(s) All results CONFIGU-...
  • Page 61: What's Inside

    DOC. MIE92093 Rev. 1.34 Page 61 of 108 3 WH AT’ S I N S IDE ? 3.1 P HYSICAL DESCRIPTION The test set is contained in an housing, to which it is fixed by means of nine screws that are located below it: see the following picture.
  • Page 62 DOC. MIE92093 Rev. 1.34 Page 62 of 108 The mechanical components are: . The front plate, on which are mounted: the VARIAC transformer, all interfacing components and the front board; . The base plate, on which are mounted all other components; .
  • Page 63 16) Converter board code PWA11401. 17) Front panel board; code PWA11398. 24) Resistors set. 26) MISU T 1000 PLUS board code PWA11402. 27) Protection fuses board, code PWA11416. The arrangement is shown in the figure and pictures below. The picture above shows the base with the auxiliary DC and AC modules removed.
  • Page 64 DOC. MIE92093 Rev. 1.34 Page 64 of 108 On the upper corner are mounted the control cards: they are: MICROT 1000 PLUS, PWA41300; CONVT 1000 PLUS, PWA11401; INTET 1000 PLUS, PWA11400. they are connected to the back- panel PWA11399, and to the front panel via three flat cables.
  • Page 65: Detailed Function Description

    DOC. MIE92093 Rev. 1.34 Page 65 of 108 Their location is shown in the drawing below. Cards are kept in place by the piece of aluminium sheet on the right, and also by the threaded bar between the boards. If a board is failing, it is necessary to remove the two screws as shown, and to un-tighten the nut of the threaded bar, as shown;...
  • Page 66: Auxiliary Transformer, Xtf10245 (4)

    DOC. MIE92093 Rev. 1.34 Page 66 of 108 VOLTAGE POWER PURPOSE GOES TO Supply of auxiliary DC voltage PWA11395 module YWA11395 (9) J800-1; 4 11.5 Auxiliary supply for auxiliary PWA11395 DC voltage module J800-2, 3 YWA11395 Low power test supply 2 WAY CONN.
  • Page 67 DOC. MIE92093 Rev. 1.34 Page 67 of 108 Positive voltage for the PWA11396 buffer of AC voltage J2-4, J1-6 board PWA11396 (10) Negative voltage for the PWA11396 buffer of AC voltage J1- 4, 5 board PWA11396 (10) 11.5 x 2 1.5 x 2 I measurement circuit PWA11398...
  • Page 68: Main Output Transformer, Xtf10345 (5)

    DOC. MIE92093 Rev. 1.34 Page 68 of 108 3.2.3 Main output transformer, XTF10345 (5) The main output transformer (5) has the following secondary windings. For connections, see the schematic at the end of this manual. VOLTAGE POWER PURPOSE 225 + Main DC voltage Main AC voltage (same common) 10 A output (60 s maximum)
  • Page 69: Conv T1000 Board Pwa11401 (16)

    CONV type: see dip 4 T 1000 PLUS at 15 Hz . ON = 15 Hz; OFF = 40..70 Hz T 1000 PLUS power. ON = more power; OFF = normal. CONV type: dip sw4 ON , dip sw1 OFF = CONV...
  • Page 70: Inte On-Off T 1000 Board Pwa11400 (7)

    DOC. MIE92093 Rev. 1.34 Page 70 of 108 3.2.7 INTE ON-OFF T 1000 board PWA11400 (7)  TASK 1: to detect the START and STOP inputs closed – open situation via two isolated constant current generator circuits, with wet/dry selection and opto-isolated logic output. ...
  • Page 71: Connectors Summary

    DOC. MIE92093 Rev. 1.34 Page 71 of 108 By the side of the boards support is mounted a board with a set of protection fuses. They protect the following modules: . Auxiliary DC voltage supply: timed 0.25 A; . Display backlight: timed 1.6 A; .
  • Page 72 DOC. MIE92093 Rev. 1.34 Page 72 of 108 J812: MAIN BOARD (17) VOLTAGE 11,5 V REF. (1-3) 11,5 V 11,5 V REF. (4-6) 11,5 V TRANSFO (4) J815: MAIN BOARD (17) VOLTAGE 8 V REF. (1) TRANSFO (4) - J817: MAIN BOARD (17) VOLTAGE 7 V REF.
  • Page 73: The Hell, It Doesn't Work

    Murphy’s law, when it was most necessary. We at ISA do our best efforts to filter the so-called infant mortality of electronic components prior to delivery of all our test sets;...
  • Page 74: Error Messages

    DOC. MIE92093 Rev. 1.34 Page 74 of 108 4.2 E RROR MESSAGES The test set performs a number of tests at power-on and during the generation. The following table lists all the messages, and the corrective action. A) TESTS AT POWER-ON (and runtime) ERROR MESSAGE CONSEQUENCE CORRECTIVE...
  • Page 75 DOC. MIE92093 Rev. 1.34 Page 75 of 108 B) RUNTIME TESTS ERROR CONSEQUENCE CORRECTIVE MESSAGE ACTION 60 VA IAC supply The test lasted Wait until the time too long; tests message disappears are blocked for the specified duration 300 VA IAC The test lasted Wait until the supply time...
  • Page 76: Trouble Shooting

    DOC. MIE92093 Rev. 1.34 Page 76 of 108 disappears, the test set informs you about the residual time OVERTEMP Generation is The main POWER stopped transformer TRANSFORMER (NOTE 1) temperature is too high because of heavy loads or very long test duration: wait until it cools down.
  • Page 77 When in the correction you find “see text”, please proceed with the following paragraphs. SYMPTOM POSSIBLE CAUSE CORRECTION Problems after Heavy shock Open T 1000 PLUS transport and check for loose boards or connections At power-on does Mains supply fuse...
  • Page 78 DOC. MIE92093 Rev. 1.34 Page 78 of 108 Measurements are The microprocessor Go to the completely false has lost its calibration correction factors procedure and repeat it. The display is dark Protection fuse See text blown Unstable auxiliary Fault of AC voltage See text AC voltage generation circuit...
  • Page 79: Auxiliary Supplies

    DOC. MIE92093 Rev. 1.34 Page 79 of 108 4.3.1 Auxiliary supplies On the edge of the CONV board are mounted a number of ring- shaped test points. They carry the key auxiliary supplies, which operate the test set. The arrangement of the test points is the following.
  • Page 80: Auxiliary Dc Voltage Fault

    DOC. MIE92093 Rev. 1.34 Page 80 of 108 . three connectors from the DC voltage generator: two bigger, white, one on the front, one on the rear, plus one flat cable; . Five connectors from the AC voltage amplifier: four white, mounted on top of the board, plus a flat cable.
  • Page 81 DOC. MIE92093 Rev. 1.34 Page 81 of 108 4 wires with the same colour 4 wires with different colours: they come from the transformer If the potentiometer was not the problem and fuse (19) is OK, it is possible that the fuse on the fuses board is blown, or that one of the fuses mounted on the board has blown up.
  • Page 82 DOC. MIE92093 Rev. 1.34 Page 82 of 108 F 2 A F 3.15 A The last test is to remove the connector J800 coming from the big round power transformer, and located on the rear of the module. The connector has 4 pins: the voltage between pins 1 and 4 should be 155 V AC nominal;...
  • Page 83: Auxiliary Aac Voltage Fault

    DOC. MIE92093 Rev. 1.34 Page 83 of 108 4.5 A UXILIARY VOLTAGE FAULT Possible modes for AC voltage fault are: . Over-load message at power-on, that cannot be re-set; . Over-load message even with a minimum load; . No AC voltage output. In such instances, the most likely cause of fault is the amplifier (10), PWA11396.
  • Page 84: N O Output From The Main Current And Voltage

    DOC. MIE92093 Rev. 1.34 Page 84 of 108 PIN 1 PIN 2 . Verify the signals as summarized in the following table. SIGNAL 0 Vin Range + 12 V meas 0-2 V meas 80 V H = 260 10 V L = 130 SIGNAL Range sel...
  • Page 85: Does Not Measure The Main Current

    DOC. MIE92093 Rev. 1.34 Page 85 of 108 follow the instructions above for the opening of T 1000 PLUS and board replacement. 4.7 D OES NOT MEASURE THE MAIN CURRENT Remove T 1000 PLUS from the container. Just below the output sockets there are two yellow wires going to the metering current transformer.
  • Page 86 DOC. MIE92093 Rev. 1.34 Page 86 of 108 . First of all, verify that the two-way, AMP type white connector is correctly fit, and that wires are not loose. . Next, try to re-solder the four connecting pins of the converter to the main board: we once had a case of poor soldering.
  • Page 87 . If everything is OK, there is also the possibility that the MICR board is faulty: replace it. . If changing the MICR board the fault is still there, it is necessary to return the test set to ISA for the replacement of the display, that involves a difficult labor.
  • Page 88: The Ac Voltage Measurement Is Not Stable

    DOC. MIE92093 Rev. 1.34 Page 88 of 108 4.9 T VOLTAGE MEASUREMENT IS NOT STABLE First of all, check that there is no unplugged connector or broken wire. Next, it is necessary to understand if the problem is in the measurement or if the output is actually unstable.
  • Page 89 INTE ON-OFF board (7), that is located in the boards pack. The replacement is very fast: just follow the instructions above for the opening of T 1000 PLUS and board replacement. If the replacement of the INTE ON-OFF board is not solving the problem, check the following.
  • Page 90: Problems During Upgrade

    If there is a doubt about the timing metered by T 1000 PLUS, please proceed as follows. . First of all, go to the CONFIG selection; settings, and select RESTORE DEFAULT: this avoids that some exotic setting is the cause of the problem.
  • Page 91: The Encoder Is Broken

    DOC. MIE92093 Rev. 1.34 Page 91 of 108 - Remove (or unlock a little) the metal bar that fixes the two aluminium columns. - Remove the two connectors placed on the INTE board. - Remove the INTE board, and then the pack MICRO + CONV. - Open the pack MICRO + CONV.
  • Page 92: The Thermal Alarm Does Not Disappear

    DOC. MIE92093 Rev. 1.34 Page 92 of 108 . Unsolder the first wire, and solder it to the new encoder, in the same position: with this procedure you are sure not to exchange wires. . At the end of soldering, mount back the new encoder, power-on and check that it is correctly operational.
  • Page 93: The Fault Cannot Be Fixed

    DOC. MIE92093 Rev. 1.34 Page 93 of 108 . The fuse to the right protects the auxiliary contact no. 1. Replace the fuse and verify again. . If the fuse is not the problem, it is possible, even if quite difficult, to unsolder the relay and to replace it.
  • Page 94 DOC. MIE92093 Rev. 1.34 Page 94 of 108 INSTRUMENT RETURN FORM DATE ________________________________________ AGENT _______________________________________ COUNTRY _____________________________________ TYPE OF INSTRUMENT __________________________ SERIAL NO. ___________________________________ INSTRUMENT RETURNED FOR: CALIBRATION ____ REPAIR ____ In case of repair, please specify the following. DATE OF FAULT _________________________________ REPORTED BY E-MAIL, PHONE ______________________ COMPANY ______________________________________ USER’S REFERENCE _______________________________...
  • Page 95 DOC. MIE92093 Rev. 1.34 Page 95 of 108 HOW DID IT OCCUR _______________________________________________ _______________________________________________ _______________________________________________ _______________________________________________ _______________________________________________ _______________________________________________ _______________________________________________ _______________________________________________ _______________________________________________ LOCAL ANALYSIS OR ATTEMPTS TO REPAIR _______________________________________________ _______________________________________________ _______________________________________________ _______________________________________________ _______________________________________________ _______________________________________________ _______________________________________________ _______________________________________________ RECOMMANDATIONS AND NOTES _______________________________________________ _______________________________________________ _______________________________________________ _______________________________________________ _______________________________________________...
  • Page 96: Calibration

    4.16 C ALIBRATION 4.16.1 Introduction T 1000 PLUS does not need to be calibrated, as metering circuits employ high stability components. It is suggested to check the unit every 2 years, by comparing T 1000 PLUS measurements to external meters. Tests should be performed with an high accuracy multi-meter, that should guarantee a maximum AC measurement error of 0.1%, both for...
  • Page 97: T 1000 Plus Output Calibration

    DOC. MIE92093 Rev. 1.34 Page 97 of 108 4.16.3 T 1000 PLUS output calibration Calibrations are divided between external measurements and internal generators. Starting from an internal generator, for instance Main IAC, the following window is opened. The three current outputs are displayed; for each output are...
  • Page 98 . Connect the meter to the output to be calibrated; then, press > to generate the output. The external meter measures the output; T 1000 PLUS displays its measurement. Adjust the current until the external meter displays a value within 5% of the nominal calibration value.
  • Page 99: T 1000 Plus External Meas. Calibration

    The auxiliary DC voltage output window is the following. . In this instance, DON’T MODIFY THE OFFSET CALIBRATION: the minimum value for the DC voltage is not zero. Just calibrate the measurement. 4.16.4 T 1000 PLUS external meas. calibration...
  • Page 100: Phase Angle Measurement Calibration

    DOC. MIE92093 Rev. 1.34 Page 100 of proceed with external measurements calibration. For this operation, IT IS NECESSARY TO HAVE AN EXTERNAL SOURCE, SUCH AS A DRTS TEST SET: T1000 ittself cannot be used to source the inputs. . The window for the external IAC measurement is the following one.
  • Page 101 DOC. MIE92093 Rev. 1.34 Page 101 of 1) Angle Auxiliary AC Voltage (reference) – Main AC current. Once the selection is performed, the following window is opened. The calibration has to be performed for each voltage and current range: nine calibrations in all. Connect the auxiliary AC voltage output to input 1 of the reference phase meter, and the main current output to input 2 of the reference phase meter.
  • Page 102 DOC. MIE92093 Rev. 1.34 Page 102 of A sockets: the auxiliary AC voltage will be the same for all ranges. 2) Angle Auxiliary AC Voltage (reference) – Main Voltage. Once the selection is performed, the following window is opened. Connect the auxiliary AC voltage output to input 1 of the reference phase meter, and the main AC voltage output to input 2 of the reference phase meter.
  • Page 103 DOC. MIE92093 Rev. 1.34 Page 103 of resistor set. Press the key >, and adjust 1 A for the first calibration, and 5 A for the second one. The procedure is the same as above. 4) Angle Auxiliary AC Voltage (reference) – External Voltage.
  • Page 104 DOC. MIE92093 Rev. 1.34 Page 104 of common With this connection, the voltage to VEXT is about 10 V when the input is 65 V: this is to be used for the angle calibration of the 20 V External input range. For other ranges, the partitioning is not necessary.
  • Page 105: Appendix 1 Spare Parts List

    DOC. MIE92093 Rev. 1.34 Page 105 of AP PE N DI X 1 S P AR E P A R T S L IS T The following is the suggested list of spare parts for a duration of five years and for up to five test sets. The reference is made to the list of components.
  • Page 106 ( 9 ) XTF10330 YWA 11395 VC C 0.5 QUADRI L= cm J800- 1 J818- 1 BI AN CO 0.5 QUADRI L= cm J800- 4 J818- 2 ( 17 ) 0.5 QUADRI L= cm J800- 2 J818- 3 PWA 11398 BP A PP O G G I O 11 0V 0.5 QUADRI L= cm J800- 3...
  • Page 108: Appendix 2: Menu Selections Flux Diagram

    APPENDIX 2: MENU SELECTIONS FLUX DIAGRAM...

Table of Contents