10. Problems and solutions

PROBLEM
Actuator does not work
Batteries do not charge
The actuator stops

11. Maintenance

All SUNRISE MEDICAL products are designed for minimal maintenance, but some regular safety checks are necessary.
SUNRISE MEDICAL recommends that the following checks be made every day, before the hoist is used.
• Ensure that the front castors of the hoist move normally.
• Ensure that the spreader bar rotated and rocks freely.
• Check that the spreader bar is securely attached to the boom.
• Ensure that the legs of the hoist open and close normally.
• Ensure that the hand control unit is being used properly.
• Check that the emergency stop button works properly.
• Check for wear on the hooks. Do not use slings which show signs of wear or damage in the lifting straps or sling body.
SUNRISE MEDICAL recommends that the following parts of hoists be serviced by an authorised technician at least once a year:
• SPREADER BAR
- Check that the spreader bar rotates and rocks freely
- Check that the spreader bar is securely attached to the hoist boom
• BOOM
- Check the coupling between boom and mast
- Check that the boom rotates fully on its bearing
• MAST
- Check that the mast is properly fitted into its plinth
- Check the electrical actuator assembly, both ends fixed with nut and bolt.
• BATTERY CASE
- Check that the emergency stop button works properly
- Check the plug of the hand control unit and actuator
• LEG ADJUSTMENT: Check legs open and close smoothly.
• LEG PIVOTS
- Check that these are firmly secured
- Check that the legs pivot freely
- If they are stiff, lubricate them with mineral oil
- Check for excessive play
• FRONT CASTORS
- Check that these are firmly secured to the legs
- Check that castors and their supports rotate freely
- Clean away any threads, hairs or fluff which might have built up.
- Lubricate with mineral oil if necessary
• ELECTRIC ACTUATOR : Check that this works properly and without any abnormal noises
• BATTERIES
- Recharge regularly
- Do not allow them to become completely discharged
• CHARGE TEST : The charge test should be performed by an authorised service technician
CAUSE
Emergency button pressed
Total battery discharge protection activated
Poor connection of the control plug
Fully charged battery (green light)
Overload in the hoist
Batteries discharged (sound warning when pressing
the hand control: protection against total discharge
activated)
SOLUTION
Turn it in the sense of the arrows
Charge the batteries
Reinsert the control plug
Verify power source
Reduce the load on the actuator
Charge batteries