Technical Data
Measuring principle
Measuring system
Measured variable
Measuring range
Type of input
Input
Resistance thermometer
(RTD)
to IEC 751
(α = 0.00385)
to DIN 43760
(α = 0.006180)
Resistance transmitter
20
10
Technical Data
10.1
Function and system design
Electronic monitoring and conversion of input signals in industrial temperature
measurement.
The temperature DIN rail transmitter is a two wire transmitter with an analog output. It
has measurement input for resistance thermometers (RTD) in 2-, 3- or 4-wire
connection, thermocouples and voltage transmitters. Setting up of the unit is done using
the configuration set for PC.
10.2
Input
Temperature (temperature linear transmission behavior), resistance and voltage
The transmitter monitors different measuring ranges depending on the sensor
connection and input signals.
Designation
Pt100
Pt500
Pt1000
Ni100
Ni500
Ni1000
•
Connection type: 2-, 3- or 4-wire connection
•
Software compensation of cable resistance possible in the 2-wire system
(0 to 20 Ω)
•
Sensor cable resistance max. 11 Ω per cable in the 3- and 4-wire system
•
Sensor current: ≤ 0.6 mA
Resistance Ω
Temperature DIN rail transmitter
Measuring range limits
-328 to 1562 °F (-200 to 850 °C)
-328 to 482 °F (-200 to 250 °C)
-328 to 482 °F (-200 to 250 °C)
-76 to 356 °F (-60 to 180 °C)
-76 to 302 °F (-60 to 150 °C)
-76 to 302 °F (-60 to 150 °C)
10 to 400 Ω
10 to 2000 Ω
Min. span
18 °F (10 °C)
18 °F (10 °C)
18 °F (10 °C)
18 °F (10 °C)
18 °F (10 °C)
18 °F (10 °C)
10 Ω
100 Ω
Endress+Hauser