Microlife WatchB Home T. Instruction Manual page 41

Table of Contents

Advertisement

Guidance and manufacturer's declaration – electromagnetic immunity –for all EQUIPMENT and SYSTEMS
Guidance and manufacturer´s declaration – electromagnetic immunity
The Model WatchBP Home T(BP3MX1-3T) are intended for use in the electromagnetic environment
specified below. The customer or the user of the Model WatchBP Home T(BP3MX1-3T) should
assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrostatic
transient / burst
IEC 61000-4-4
Surge
IEC 61000-4-5
IEC 60601 test level Compliance level
± 8 kV contact
±2 kV, ±4 kV, ±8 kV, ±2 kV, ±4 kV, ±8 kV,
±15 kV air
± 2 kV for power
supply lines 100
kHz repetition
frequency± 1 kV for
input/output lines
± 0.5 kV, ± 1 kV
differential mode line-
line
± 8 kV contact
±15 kV air
± 2 kV for power
supply lines100 kHz of a typical commercial or hospital
repetition frequency environment.
± 0.5 kV, ± 1 kV
differential mode line-
line
Electromagnetic environment -guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Mains power quality should be that
Mains power quality should be that of a
typical commercial or hospital
environment.
BT
41
EN