Download Print this page

Agilent Technologies G4319A Installation Manual page 8

Tcd-to-fid series connection on a 6820 gc, accessory

Advertisement

Installing the TCD-to-FID Series Connector
TCD column or make up
gas adapter fitting
TCD vent tube
1/8-inch stainless tubing nut
1/8-inch stainless back ferrule
1/8-inch stainless front ferrule
TCD-to-FID jumper tube
8
1 Working inside the oven, and following proper swaging
procedures described in the 6820 Getting Started manual,
use the 1/8-inch stainless tubing nut, back ferrule, and front
ferrule to connect the TCD-to-FID jumper tube to the TCD
vent tube. Use two wrenches against each other to avoid
twisting the vent tube.
2 Use the 1/4-inch stainless tubing nut and Vespel ferrule to
install the 1/4- to 1/8-inch adapter into the FID.
3 Again, following proper swaging procedures, use the 1/8-inch
brass tubing nut, back ferrule, and front ferrule to connect
the TCD-to-FID jumper tube to the FID 1/4- to 1/8-inch
adapter.
FID column or column
adapter fitting
1/4-inch Vespel ferrule
1/4-inch stainless tubing nut
FID 1/4- to 1/8-inch
column adapter
1/8-inch brass front ferrule
1/8-inch brass back ferrule
1/8-inch brass tubing nut
Installation Guide: TCD-to-FID Series Connection

Advertisement

loading