Accuracy Verification Procedure
The procedure verifies the Energy Logger accuracy at ambient temperature
23 °C ±2 K (intrinsic error).
A complete accuracy verification of the Fluke 1730 consists of:
•
Voltage Measurement
•
Current Measurement
•
AUX Measurement
•
Optional Flexi or Current Verification
Voltage Measurement
1. Select the basic instrument setup required for all verifications, see page 18.
2. Connect the VL1730 "N" lead to the calibrator NORMAL LO.
3. The 1730 must be on battery power with ≥50 % charge.
4. Connect the calibrator NORMAL V output to the VL1730 L1+L2+L3 leads.
Sequentially set the calibrator to the voltages indicated in Table 9 and check
5.
that the Energy Logger reading is between the limits.
Do this for all ranges indicated in Table 9:
6.
•
set the calibrator to supply a 57.0 Hz sine wave for all voltages
•
wait until each reading has stabilized
The spreadsheet is the first choice for readings. Readings will have more
resolution from the spreadsheet.
7. Push to select the Energy Logger voltage display.
Nominal Voltage
(Range)
1000
1000
1000
1000
8. When you are done, set the calibrator to Standby.
Table 9. Voltage Verification
Calibrator voltage
(57 Hz sine wave)
10
100
500
1000
Accuracy Verification Procedure
Minimum Reading
± (1% + 0.1%)
9.9
99.7
498.9
998
Energy Analyzer
Maximum Reading
10.1
100.3
501.1
1002
19