4. Maintenance
4.1 General maintenance procedures
The performance and safety of the instrument will be kept if the following recommendations for care and
maintenance are observed:
The instrument must go through at least one annual overhaul, to ensure that the acoustical, electrical
•
and mechanical properties are correct. This should be made by an authorised workshop in order to
guaranty proper service and repair as Interacoustics provides the necessary circuit diagrams etc. to
these repair shops.
To ensure that the reliability of the instrument is kept, it is recommended that the operator at short
•
intervals, for instance once a day, perform a test on a person with known data. This person could be
the operator him/herself.
•
After each examination of a patient, it should be ensured that there is no contamination on the parts
in connection with the patient. General precautions must be observed in order to avoid that disease
from one patient is conducted to others. If ear cushions or ear tips are contaminated, it is strongly
recommended to remove them from the transducer before they are cleaned. By frequent cleaning
water should be used, but by severe contamination it may be necessary to use a disinfectant. The
use of organic solvents and aromatic oils must be avoided.
4.2 How to clean Interacoustics products
Before cleaning always switch off and disconnect from the power supply
•
•
Use a soft cloth lightly dampened with cleaning solution to clean all exposed surfaces
Do not allow liquid to come in contact with the metal parts inside the earphones/headphones
•
•
Do not autoclave, sterilize or immerse the instrument or accessory in any fluid
Do not use hard or pointed objects to clean any part of the instrument or accessory
•
Do not let parts that have been in contact with fluids dry before cleaning
•
Rubber ear-tips or foam ear-tips are single use components
•
Ensure that isopropyl alcohol does not come into contact with any screens on the instruments
•
•
Ensure that isopropyl alcohol does not come into contact with any silicone tubes or rubber parts
Recommended cleaning and disinfection solutions:
Warm water with mild, nonabrasive cleaning solution (soap)
•
•
Normal hospital bactericides
70% isopropyl alcohol only on hard cover surfaces
•
Procedure
•
Clean the instrument by wiping outer case with a lint free cloth lightly dampened in cleaning solution
Clean cushions and patient hand switch and other parts with a lint free cloth lightly dampened in
•
cleaning solution
Make sure not to get moisture in the speaker portion of the earphones and similar parts
•
D-0101246-M – 2020/05
Callisto™ - Instructions for Use - US