Callisto™ - Instructions for Use - US
Current Protocol is listed in the lower left hand corner.
The
indicates the test is a part of an automatic test flow
(Auto Run). When pressing START all tests with the tick
mark will be performed.
test.
Upon performing a test the system automatically jumps to
the next one in the test flow.
been measured.
Colour indication shows the colour selected for each curve.
Test protocols can be created and adjusted in the HIT440
Setup.
Start/Stop button initiates and ends all tests.
Note that after pressing START the text on the button will
change to STOP.
The Graph shows measured HIT curves. The X
axis shows the frequency and the Y axis shows
ouput or gain, depending on which measurement
was done.
Measurement type is printed above the graph
together with a right/left indication. In this example
the OSPL90 is displayed for the left ear.
Change the input level using the slider on the right
hand side.
NOTE: for the industry standard protocols (ANSI
and IEC), the input level is dictated by the standard
and cannot be changed.
Scroll graph up/down on the left hand side allows
for scrolling the graph up or down ensuring that the
curve is always visible in the middle of the screen.
If you wish to perform one test only,
mark it using the mouse by clicking on
it. Then right click at select Run this
indicates that a curve has