Interacoustics Callisto AC440 Instructions For Use Manual page 33

Hide thumbs Also See for Callisto AC440:
Table of Contents

Advertisement

1
HF requires an additional feature for the AC440. If not purchased, the button is greyed out.
2
MF requires additional feature for the AC440. If not purchased, the button is greyed out.
D-0101246-M – 2020/05
Callisto™ - Instructions for Use - US
Fold an area so that it only shows the label or the buttons of that area.
Unfold an area so that all buttons and labels are visible.
Show/hide areas can be found by right mouse clicking on one of the
areas. The visibility of the different areas as well as the space that they
take on the screen is locally saved to the examiner.
List of Defined Protocols allows for selecting a test protocol for the
current test session. Right mouse click on a protocol allows the current
examiner to set or deselect a default startup protocol. Please refer to the
"Additional Information" document for more information about protocols
and protocol setup.
Temporary Setup allows for making temporary changes to the selected
protocol. The changes will be valid for the current session only. After
making the changes and returning to the main screen, the name of the
protocol will be followed by an asterisk (*).
List of historical sessions accesses historical sessions for comparison
purposes. The audiogram of the selected session, indicated by the
orange background, is shown in colours as defined by the used symbol
set. All other audiograms that are selected by check marks show on
screen in the colours as indicated by the text colour of the date and time
stamp. Note that this listing can be resized by dragging the double lines
up or down.
Go to Current Session brings you back to the current session.
High Frequencies
1
show frequencies on the audiogram (up to 16 kHz for
the Callisto™). However, you will only be able to test in the frequency
range the high frequency headset is calibrated for.
High Frequency Zoom activates high frequency testing and zooms in on
the high frequency range. Please refer to the Callisto™ "Additional
Information" document for more information on HF Audiometry
Toggle between Single and Dual Audiogram button toggles between
viewing the audiogram as one combined graph or two separate graphs.
Multi Frequency
2
activates Multi Frequency testing. The frequency steps
can be adjusted in the AC440 setup.
Synchronize channels locks the two channels together. This function
may be used to perform synchronous masking.