Bühler technologies CPsingle X1 Installation And Operation Instructions Manual

Peristaltic

Advertisement

Quick Links

Analysentechnik
Peristaltic pumps
CPsingle X1, CPdouble X1
Installation and Operation Instructions
Original instructions
BE450026
Bühler Technologies GmbH, Harkortstr. 29, D-40880 Ratingen
09/2018
Tel. +49 (0) 21 02 / 49 89-0, Fax: +49 (0) 21 02 / 49 89-20
E-Mail: analyse@buehler-technologies.com
Internet: www.buehler-technologies.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the CPsingle X1 and is the answer not in the manual?

Questions and answers

Summary of Contents for Bühler technologies CPsingle X1

  • Page 1 Analysentechnik Peristaltic pumps CPsingle X1, CPdouble X1 Installation and Operation Instructions Original instructions BE450026 Bühler Technologies GmbH, Harkortstr. 29, D-40880 Ratingen 09/2018 Tel. +49 (0) 21 02 / 49 89-0, Fax: +49 (0) 21 02 / 49 89-20 E-Mail: analyse@buehler-technologies.com...
  • Page 2 Bühler Technologies GmbH, Harkortstr. 29, D-40880 Ratingen Tel. +49 (0) 21 02 / 49 89-0, Fax: +49 (0) 21 02 / 49 89-20 Internet: www.buehler-technologies.com E-Mail: analyse@buehler-technologies.com Read this instruction carefully prior to installation and/or use. Pay at- tention particularly to all advises and safety instructions to prevent in- juries.
  • Page 3: Table Of Contents

    CPsingle X1, CPdouble X1 Contents Introduction.............................................  2 Intended use .........................................  2 Scope of delivery ........................................ 2 Type plate .......................................... 2 Peristaltic pump ordering information................................  2 Safety instructions.........................................  3 Important advice .........................................  3 General hazard warnings .................................... 4 Transport and storage ........................................ 5 4 Installation and connection ...................................... 6...
  • Page 4: Introduction

    CPsingle X1, CPdouble X1 1 Introduction 1.1 Intended use This unit is intended to discharge condensate from cooled process fluids. The temperature of these mediums is approx. 5 °C. These operating instructions apply to pressurised encapsulation with surface-cooled electric motors for low voltage, degree of protection IP40, IP54, IP64, IP65 EN 60034-5:2001+A1:2007.
  • Page 5: Safety Instructions

    CPsingle X1, CPdouble X1 2 Safety instructions 2.1 Important advice This unit may only be used if: – The product is being used under the conditions described in the operating- and system instructions, used according to the nameplate and for applications for which it is intended. Any unauthorized modifications of the device will void the warranty provided by Bühler Technologies GmbH,...
  • Page 6: General Hazard Warnings

    CPsingle X1, CPdouble X1 2.2 General hazard warnings The equipment must be installed by a professional familiar with the safety requirements and risks. Be sure to observe the safety regulations and generally applicable rules of technology relevant for the installation site. Prevent malfunctions and avoid personal injuries and property damage.
  • Page 7: Transport And Storage

    CPsingle X1, CPdouble X1 3 Transport and storage The products should be transported only in its original packaging or a suitable replacement. When not in use, protect the equipment against moisture and heat. Keep it in a covered, dry and dust-free room.
  • Page 8: Installation And Connection

    CPsingle X1, CPdouble X1 4 Installation and connection 4.1 Installation site requirements Be sure to maintain the approved ambient temperature. Please also note the technical data of the add-on gas cooler. When mounting to a subframe, it screws directly onto the cooler housing frame.
  • Page 9 CPsingle X1, CPdouble X1 Be sure to observe any varying information in the rating plate. The conditions at the site must correspond with all rating plate information. Protective measures against prohibited heating Unless otherwise certified, the electric motor may only be used for continuous operation, not frequent restarting without signi- ficant start-up warming.
  • Page 10: Operation And Control

    CPsingle X1, CPdouble X1 5 Operation and control NOTICE The device must not be operated beyond its specifications. The pump does not have a power switch. It starts running as soon as the power supply is turned on. NOTICE Installing peristaltic pumps CPsingle / CPdouble limits the maximum permissible oper- ating pressure in the system! Operating pressure ≤...
  • Page 11: Maintenance

    CPsingle X1, CPdouble X1 6 Maintenance During maintenance, remember: – The equipment must be maintained by a professional familiar with the safety requirements and risks. – Only perform maintenance work described in these operating and installation instructions. – When performing maintenance of any type, observe the respective safety and operation regulations.
  • Page 12: Replacing The Hose

    CPsingle X1, CPdouble X1 6.2 Replacing the hose – Close gas supply. – Switch off device and disconnect all plugs (e.g. connector plug alarm output, supply input, etc.). – Disconnect supply and discharge tube on peristaltic pump (observe safety notes!).
  • Page 13: Changing The Pump Head/Rotor

    CPsingle X1, CPdouble X1 6.3 Changing the pump head/rotor Single and double pump heads fix the same. The individual parts vary based on the version. Attention! The cylindrical stud of the rotor main body always faces forward Attention! The cylindrical stud of the...
  • Page 14: Service And Repair

    CPsingle X1, CPdouble X1 7 Service and repair This chapter contains information on troubleshooting and correction should an error occur during operation. Repairs to the unit must be performed by Bühler authorised personnel. Please contact our Service Department with any questions: Tel.: +49-(0)2102-498955 or your agent...
  • Page 15: Spare Parts And Accessories

    CPsingle X1, CPdouble X1 7.2 Spare parts and accessories Spare parts and wear item requirements vary based on the specific operating conditions of the pump. Please also specify the model and serial number when ordering parts. Upgrade and expansion parts can be found in our catalog.
  • Page 16: Disposal

    CPsingle X1, CPdouble X1 8 Disposal Dispose of parts so as not to endanger the health or environment. Do not dispose of the electric motor and the pressurised encapsulation as unsorted municipal waste. If necessary, return directly to the manufacturer.
  • Page 17: Appendices

    CPsingle X1, CPdouble X1 9 Appendices 9.1 Technical data Technical Data CPsingle/CPdouble Peristaltic Pumps Nominal voltage / power input: 230 V 50/60 Hz, 0.026 A (50/60 Hz) at T = 20 °C and under load 115 V 50/60 Hz, 0.052 A (50/60 Hz) ±5 % voltage, ±2 % frequency...
  • Page 18: Version With 2 Gas Paths

    CPsingle X1, CPdouble X1 9.2.2 Version with 2 gas paths Bühler Technologies GmbH BE450026 ◦ 09/2018...
  • Page 19: Attached Documents

    CPsingle X1, CPdouble X1 10 Attached documents – Certificate IECEx IBE 17.0019 – Type Examination Certificate IBExU17ATEX1008 – Declaration of Conformity KX450016 – RMA - Decontamination Statement BE450026 ◦ 09/2018 Bühler Technologies GmbH...
  • Page 22    ...