Electromagnetic Compatibility Guidelines; Safety Guidelines For Hazardous Voltages; Verifying The System Requirements - National Instruments PXIe-4135 Getting Started Manual

Single-channel precision system source measure unit (smu)
Hide thumbs Also See for PXIe-4135:
Table of Contents

Advertisement

Where To Go Next.................................................................................................................. 17
Worldwide Support and Services............................................................................................ 17

Electromagnetic Compatibility Guidelines

This product was tested and complies with the regulatory requirements and limits for
electromagnetic compatibility (EMC) stated in the product specifications. These requirements
and limits provide reasonable protection against harmful interference when the product is
operated in the intended operational electromagnetic environment.
This product is intended for use in industrial locations. However, harmful interference may
occur in some installations, when the product is connected to a peripheral device or test object,
or if the product is used in residential or commercial areas. To minimize interference with
radio and television reception and prevent unacceptable performance degradation, install and
use this product in strict accordance with the instructions in the product documentation.
Furthermore, any changes or modifications to the product not expressly approved by National
Instruments could void your authority to operate it under your local regulatory rules.
Caution
shielded cables and accessories connected to the LO/Sense LO, Output HI, and
Sense HI connectors. Only twisted, shielded pair cable (i.e. Belden 83321E) should
be used to connect the signals of the LO/Sense LO connector. Any compatible
unshielded twisted-pair cable may be used for the safety interlock connector.
Caution
must be no longer than 3 m (10 ft).

Safety Guidelines for Hazardous Voltages

If hazardous voltages are connected to the device, take the following precautions. A hazardous
voltage is a voltage greater than 42.4 Vpk voltage or 60 VDC to earth ground.
Caution
signal voltages no greater than 250 V DC. This module can withstand up to
500 V DC impulse voltage. Do not use this module for connection to signals or for
measurements within Categories II, III, or IV. Do not connect to MAINS supply
circuits (for example, wall outlets) of 115 V AC or 230 V AC.
Caution
channel pin and the chassis ground. When operating channels in series or floating on
top of external voltage references, ensure that no terminal exceeds this rating.

Verifying the System Requirements

To use the NI-DCPower instrument driver, your system must meet certain requirements.
2 | ni.com | PXIe-4135 Getting Started Guide
To ensure the specified EMC performance, operate this product only with
To ensure the specified EMC performance, the length of all I/O cables
This module is rated for Measurement Category I. It is intended to carry
Isolation voltage ratings apply to the voltage measured between any