TS1201 PINCH TUBE ASSEMBLY REPLACEMENT
TO REPLACE PINCH TUBE ASSY.
1. Unscrew pivot Pin (a), remove Trigger (b), and
Spring (c).
2. Pull one end of Pinch Tube Assy. out (d), stretching
it until fitting clears body.
3. Now pull stretched end of Pinch Tube Assy. up (e).
Then pull out the opposite end (f) to remove.
4. To replace new Pinch Tube Assy., insert fitting of
Pinch Tube Assy. in one end of the body (g). Stretch
the opposite end over the body until fitting clears (h).
5. Now pull Pinch Tube Assy. down (i) and insert end
into the body (j). Finally, secure by pushing Pinch Tube
Assy. to the bottom inside of body (k) and re-assemble.
NOTE: After installing new Pinch Tube Assy. it may be necessary to purge air from line. Refer to
section 4.2, Number 4. If Pen Trigger can not instantly shut off adhesive flow, air is not completely
purged from dispense pen material line.
f
g
Page 15
b
c
a
k
FIGURE 8
d
e
h
i
j