e. Activate dispense option (i.e. Press Trigger on Pen, or apply air to Valve), slowly turn Air Press Knob
clockwise. Dispense all adhesive into container. Dispense solvent into container for 5 to 10 seconds after
adhesive is dispensed. Dispose of properly.
4. Purge all remaining cleaner with air.
5. Turn air pressure regulator counter clockwise and lower air pressure to zero.
6. Store adhesive dispenser with Lid open.
6.1 CLEANING/REPAIR- TS1201 PEN
1. To remove Pen from supply line:
a. Turn Air Pressure Regulator counter clockwise and lower air pressure to zero.
b. Remove needle and hold dispense Pen higher than pressure Reservoir.
c. Press Dispense Pen Trigger and drain adhesive into adhesive Bottle. If cured adhesive plugs Pinch
Tube, adhesive will not drain into adhesive Bottle. Refer to in Section 6.0, Note 3, Instructions d & e.
d. Remove supply line Tube from Dispense Pen.
NOTE: If leakage occurred in Dispense Pen, soak entire Pen in Methylene Chloride for disassembly. Clean up
uncured adhesive with Methyl Ethyl Ketone or Nitromethane. Remove cured adhesive with Nitromethane.
Acetone or Methyl Ethyl Ketone works slowly on cured adhesive.
2. Soak parts of dissembled Pen in Cyanoacrylate cleaner (Refer to page 5).
3. To replace Pinch Tube (Refer to figure 8).
6.2 CLEANING/REPAIR- TS1212 VALVE
1. To clean Valve follow Instruction in Section 6.1 Numbers 1 & 2.
2. To replace Pinch Tube (Refer to figure 9).
Page 14