Terragene Bionova® IC10/20FR Auto-reader has been ratory or hospital. The auto-reader allows selection of designed for the incubation of the following Self-Con-...
To avoid BIs from absorbing fluorescent re- sidues: avoid BIs direct contact with chemical Serial number. indicators or tapes. Direct current. Do not use this product in a manner not speci- fied by Terragene® S.A., otherwise the protec- Product denomination. IC10/20FR Page 4...
Units Output parameters DC Volts Voltage Ampers Current Terragene® recommends using a UPS instead of stabi- lizers, since the former has two functions: it stabilizes and provides a continual power source during energy cutoff. Environment operating conditions Environmental conds. Functional conds.
3-Synchronize IC10/20FR with local time using Biono- auto-reader) and make sure that the liquid has va® software available at http://www.terragene.com. completely wet the spore carrier placed at the ar/productos/infection-control/incubators/softwa- bottom of the biological indicator plastic tube.
Instructions for use 8-If the incubator detects a positive result, the red light 1. To replace the paper and the audible alarm will turn on. This will indicate roll, pull the handle of that the sterilization process to which the BI was sub- the printer’s door.
C (for control) and the date on the label. The posi- and safety regulations (see BIs instructions for use). tive control can be placed in any well of the Terragene Bionova® IC10/20FR Auto-reader. It is a good practice to use a positive control, this...
Software for Automatic Reading and Traceability of Ra- Care and cleaning pid and Super Rapid Biological Indicators. You can also download it from http://www.terragene.com.ar/pro- Clean the exterior surface of the Terragene Bionova® ductos/infection-control/incubators/software-down- IC10/20FR Auto-reader with a damp cloth moistened loads/ ?lang=en.
Instructions for use Time zone change Time setting mode To advance time, press button as many times as hours you need to advance. To delay time press Time setting functionality allows you to establish a tton as many times as hours you need to delay. specific time regardless of auto-reader current time.
Page 11
Instructions for use Figure 5 Minutes setting - Minutes setting: Ten minutes (hh:Mm) Increase Decrease Press button to turn on the blue light number 3. This indicates ten minutes setting. If any other light from PRO sector is on, press button as many times as necessary to turn on the blue light number 3.
Instructions for use Finish printing language change Printing language change mode To save changes press and hold button for three seconds. A sound will be emitted indicating This function allows you to change the auto-reader that process has been completed successfully. A tic- printing language.
Troubleshooting chart Fault Possible cause Action The auto-reader does not start Power source is not connected. Check that the power source is con- nected to the power supply and that the plug is connected to the au- to-reader. The auto-reader gives an error A BI is placed in the position at Check that every position is empty at in one position during “Auto-...
Terragene® S.A. will be product repai- ring or substitution. Limitation of liability Terragene® S.A. shall not be liable for any loss or da- mage that may result from the unsuitable use of the equipment, negligence or user’s full responsibility.
Index Constituent parts ........................3 Description of use ................................3 References ....................................4 Symbols ....................................4 Safety information ........................4 Power supply specifications .............................5 Environment operating conditions ..........................5 Serial/Batch Number ................................5 Record Table ...................................5 Instructions for use ........................6 Start-up ....................................6 Thermal paper specifications ............................7 Replacement of the paper roll ............................7 Reprinting of results ................................7 Remaining incubation time ..............................7...