Real-Time Rs-232 Output - Thermo Scientific personalDATARAM pDR-1000AN Instruction Manual

Particulate monitor
Table of Contents

Advertisement

To review the parameter values currently programmed into the personalDataRAM,
click on Get configuration.
configuration to input the new values into the personalDataRAM program.
Most operations within pDR-COM are self-evidently labeled, including fly-over
dialog boxes. In addition, instructions may be found in the On-line Help files by
selecting Help and then Contents.
The following operating/logging parameters of the personalDataRAM are selected
(edited) via the computer:
• Current date (month and day of the month)
• Current time (hour, minute and second)
• Display averaging time (1 to 60 seconds, in 1-second increments)
• Calibration factor (0.01 to 9.99, in 0.01 increments)
• Logging interval (1 to 14,400 seconds, in 1-second increments)
• Analog output full scale concentration (0.1, 0.4, 1, 4, 10, 40, 100, or 400 mg/m
• Analog output status (enabled, or disabled) (can also be selected directly through
personalDataRAM keyboard, see Section 8.2)
• Alarm level (0.001 to 400 mg/m3, in 1-µg/m
• Alarm mode (Off, Instantaneous, or STEL) (can also be selected directly through
personalDataRAM keyboard, see Section 8.2)
The serial number of the personalDataRAM is transferred automatically to the PC
and displayed on its screen.
In addition, the user can input any other identification for the instrument (up to 20
characters).
9.4

Real-Time RS-232 Output

During the RUN mode, the personalDataRAM can communicate real-time
concentration data through its serial port via the pDR-COM software package. This
software application decodes the data and displays it on the computer screen in both
graphical and tabulated form.
In order to use this output with some other application, the following information
will enable the user to decipher the encoded output signal.
The communication settings for the digital output of the personalDataRAM are:
Baud rate:
Data bits:
Stop bits:
Parity:
Flow control: Xon/Xoff
After editing the parameters, click on Set
4800 bps or 9600 bps
8
1
none
30
3
increments)
3
)