8.0

OPERATION

8.1

Start-Up

KEY
1.
ON/OFF
2.
ENTER
3.
NEXT
4.
ON/OFF
*The number following the V on the screen refers to the installed firmware version.
DISPLAY
START ZERO:ENTER
GO TO RUN: NEXT
ZEROING V2.00
CALIBRATION: OK
BACKGROUND HIGH
MALFUNCTION
START RUN: ENTER
READY: NEXT
TURN OFF PDR?
Y:ENTER N:NEXT
23
NOTES
Before starting a run with the
personalDataRAM, zero it (see
Section 6.5) and key ENTER
while the unit is exposed to
particle-free air. Alternatively,
key NEXT to go to
RUN/READY mode. If ENTER
is keyed:
Keep clean air flowing while
ZEROING is displayed* for 1.1
min., followed by one of these
screens:
or,
or,
If CALIBRATION: OK, then go
to step 3. If one of the other two
screens is displayed, consult
Section 12.0.
To start a measurement run
key ENTER (Section 8.3, step
1). To set up for a run and
scroll logging/operating
parameters, key NEXT (see
Section 8.2).
Keying ON/OFF while the unit
is operating will elicit this
message to prevent accidental
shut off.
To confirm shut down, key
ENTER. To continue operation,
key NEXT.