Thermo Scientific personalDATARAM pDR-1000AN Instruction Manual page 37

Particulate monitor
Table of Contents

Advertisement

4.
NEXT
5.
NEXT
6.
NEXT
7.
NEXT
8.
NEXT
9.
NEXT
10.
EXIT
11.
ENTER
If during a run the instrument memory is filled completely, or if all 99 tags have
been used, the run is automatically terminated and the display will indicate:
If a new run is initiated after the memory has been filled, the personalDataRAM can
be operated only as a monitor without logging. The memory must then be cleared
(see Section 7.3.2) first before logging can be enabled again.
ET 06:12:49
ST 08:18:26MAY15
3
MAX: 0.113 mg/m
T 10:08:44 MAY15
3
STEL:0.058 mg/m
T 09:59:22 MAY15
BATTERY LEFT 83%
BATTERY LEFT 83%
MEMORY LEFT 96%
ANALOG OUTPUT:
3
0 – 0.400 mg/m
3
CONC*0.047 mg/m
3
TWA 0.039 mg/m
TERMINATE RUN?
Y:ENTER N:NEXT
START RUN: ENTER
READY: NEXT
RUN TERMINATED
FULL MEMORY
26
This screen shows the elapsed
run time (ET) and the run
start time/date (ST).
This screen shows the maximum
concentration of current run
and time/date of occurrence.
This screen shows the 15-min
STEL value of the current run
and the time/date of occurrence.
or, if logging was enabled:
This screen shows the amount
of usable charge left in the
battery and, if logging has
been
enabled,
the
amount of free memory left.
This screen shows the status of
the analog signal output, and the
range, if this output has been
enabled.
The last NEXT command
returns the display to the
concentration screen.
As indicated in step 2, to end
current run, key ENTER, to
return to the Ready Mode:
This keystroke terminates the
current run and returns the
unit to the Ready Mode.