Attach second adjustment weight, wait for stability display
Press
, "LoAd 3" (max load) will be displayed
Attach third adjustment weight, wait for stability display
Press
, "PASS" will be displayed
Finally the balance will carry out a self-test, "Err19" is briefly
displayed (Ignore error message), then the balance returns
automatically to weighing mode, the weight value is
displayed and linearization has been completed
successfully.
If an error occurs, turn off on restart balance and repeat linearization.