4.1.11
Ratio A/B Test
4.1.11.1
Set the 2251A to its home state. Select Ratio A/B as the measurement mode, 50 ohms
for input A input impedance, DC input coupling for input B, external frequency
standard and set the measurement resolution to three digits.
(Command String: IP RA ALI SRS3)
4.1.11.2
Connect the 2251A to the signal generator as shown in Figure 4-8. Set the signal
generator to 50 MHz, 100 mV rms. Verify that the counters measurement result is
50000 ± 10.
4.1.11.3
Disconnect the test equipment.
4-10
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Figure 4-8, Ratio A / B Test Configuration