Performing The Measurement; Barometric Pressure Compensation; Probe Ldo70 Maintenance; Replacement Of Luminophore - XS Vio OXY 70 Instruction Manual

Hide thumbs Also See for Vio OXY 70:
Table of Contents

Advertisement

Dissolved O
Saturation expressed in %
2
Dissolved
O
mg/l = ppm
Barometric pressure
During the measurement press the button
measurement.

• Performing the measurement

Remove the protective cap of electrode, rinse it with distilled water; dab it with paper towel and dip in the
solution to analyse. Gently stir and wait until the value stability, when the icon
take the reading.

• Barometric pressure compensation

Since the measurement of the partial pressure of the dissolved oxygen is related to the barometric pressure,
this instrument is able to compensate each variation, thanks to the integrated barometric sensor.
In order to view the barometric pressure measured by the instrument, press the key
measurement and scroll through the measurement parameters: %O

Probe LDO70 maintenance

If the instrument does not calibrate or the reading does not stabilize, it is necessary to perform a
maintenance of the probe in the following way:
1) Check that the luminophore is clean, in case wash it with water.
2) The luminophore must be intact, undamaged, and without holes.
3) Unscrew the luminophore and check that the inside is dry, free from condensation and infiltration.
If there is condensation or infiltration, check the integrity of the O-ring seal of the luminophore
and,
Dry thoroughly with a paper towel and screw the luminophore ensuring an airtight seal.
After the maintenance, make a new Calibration in air at 100%.
Note: if even after maintenance the probe does not calibrate, then replace the luminophore with a new one.

• Replacement of luminophore

The efficiency of the luminophore decreases with wear to the point that the instrument is no longer able to
calibrate itself, in this case it is necessary to replace it. To replace the luminophore proceed as follows:
1) Unscrew the luminophore;
2) Check the integrity of the inside of the probe;
3) Replace the o-ring with a new one supplied with a new luminophore;
4) Screw the new luminophore ensuring an airtight seal.
Concentration
expressed
2
if
in
mg/l,
, in order to change the unit of
↔ mg/l ↔ mbar.
2
necessary,
22
corresponding
to
appears on the display,
during
replace
ppm
it.