Wenden Sie sich an ihren Händler, ob Sie die mehrsprachige digitale Kopie erhalten möchten. Français: Ce manuel a été imprimé seulement en Italien et en Anglais pour réduire les déchets. Pour la version digitale multilingue veuillez contacter votre revendeur. UM Temp74 PT100 Vio EN rev.1 19.05.2023...
Pay particular attention to the technical specifications in the table INSTRUMENTAL CHARACTERISTICS / TECHNICAL DATA; any other use outside these is considered unauthorised. This instrument left the factory in perfect technical and safety condition. UM Temp74 PT100 Vio EN rev.1 19.05.2023...
Data logger Connection to PC Date and time Measuring channel Battery level 37,05 °C Measured value Stability indicator Sensor ID 123456789 Minimum value 12,2 °C 37,05 °C Maximum value Recorded Values HOLD function active UM Temp74 PT100 Vio EN rev.1 19.05.2023...
Directional key DOWN / RM OK / Setup button to scroll through menus and values in to confirm the chosen values and the Setup and manage data retrieval enter Setup from measurement mode UM Temp74 PT100 Vio EN rev.1 19.05.2023...
Switch the system on by pressing the button. They will be displayed: • device model and software; • offset of the 4 channels; • temperature measurement screen; • to switch off the instrument, press the button in measurement mode. UM Temp74 PT100 Vio EN rev.1 19.05.2023...
Number of data stored in data logger mode on instrument memory Press the arrow buttons to change the parameter or value on the display Instrument connected to the DataLink+ software Reference channel Minimum and maximum values Measuring channel displayed UM Temp74 PT100 Vio EN rev.1 19.05.2023...
• A code with a maximum length of 10 digits can be entered. Note : If you select a channel with no probe connected, the display will show the error message: "CONNECT THE PROBE", accompanied by the icon flashing UM Temp74 PT100 Vio EN rev.1 19.05.2023...
• Nor: with variations less than or equal to 2 hundredths of a degree, the stability icon remains fixed and indicates the stable value. • High: with variations less than or equal to 1 hundredth of a degree, the stability icon remains fixed and indicates the stable value. UM Temp74 PT100 Vio EN rev.1 19.05.2023...
Automatic logger - Setting the Automatic Data Logger First set the acquisition interval (hh:mm:ss) then the delay (hh:mm:ss). Use the arrows to change the flashing value. Press to confirm and move to the next value. UM Temp74 PT100 Vio EN rev.1 19.05.2023...
°C - °F °C LANGUAGE Ita - Eng - Fra - Deu - Esp AUTO SWITCH-OFF No - Yes, after 20 minutes SOUND No - Alarm only - Yes PASSWORD RESET SETTINGS No - YES UM Temp74 PT100 Vio EN rev.1 19.05.2023...
: With an automatic or programmed data-logger set, the auto-off is automatically deactivated. • Sound Access this menu to set whether the sound should always be on (every time a button is pressed), always off or only activate when the instrument detects an alarm. UM Temp74 PT100 Vio EN rev.1 19.05.2023...
Press and hold for 3 seconds to unlock the reading. Other functions • Automatic switch-off The instrument switches off automatically after 20 minutes of inactivity. To disable auto power off, go to Setup Settings Auto Power Off and select No. UM Temp74 PT100 Vio EN rev.1 19.05.2023...
• Only use the USB cable supplied with each instrument. • Start the programme and then switch on the instrument. • Wait for the connection to be established (connection data are shown at the bottom left of the display). UM Temp74 PT100 Vio EN rev.1 19.05.2023...
. Disposal This equipment is subject to regulations for electronic devices. Dispose of in accordance with existing local regulations. UM Temp74 PT100 Vio EN rev.1 19.05.2023...
Page 19
XS Instruments Via della Meccanica n.25 41012 Carpi (MO) ITALY Tel.+39059.653274 UM Temp74 PT100 Vio EN rev.1 19.05.2023...