Message on display screen: PUSH RUN TO FEED
This alarm will sound after 2 minutes unless the word RUN is on the display indicating
that the pump is running.
Solution: Press RUN/PAUSE to stop alarm; press RUN/PAUSE again to resume feed.
Listen for 3 beeps and look for RUN on the display; the pump is running.
Message on display screen: LOW BATT with distinctive alarm tone
Solution: Plug pump into charger
Message on display screen: ER##, ERRA, or ERRZ and a continuous alarm tone
Make sure pump door is closed tightly.
Press and hold ON/OFF until pump shuts down.
Restart pump by holding ON/OFF BUTTON for 3 seconds, then verify pump settings. If
settings are correct, press RUN/PAUSE.
Listen for 3 beeps and look for RUN on the display; the pump is running.
NOTE: An ER##/ERRA/ERRZ may ERASE the rate, dose, and timing of the food
being delivered. If you see 0.1 on the display when you restart the pump, contact the
parent or caregiver.
If the pump DOES NOT RESTART or you are unsure of the correct pump settings,
contact the parent or caregiver at the phone number below.