Moog EnteraLite Infinity Field Manual page 10

Feeding pump
Hide thumbs Also See for EnteraLite Infinity:

Advertisement

Message on display screen: LOAD SET
Problem: Feed set is not loaded into pump.
Solution: Load feed set into pump.
Press RUN/PAUSE.
Open DOOR of pump by pushing on the tab above the display screen.
Stretch TEAL TUBING around the outside of the black WHEEL.
Fit PLASTIC CASSETTE into SLOTS.
Check that tubing is not obstructing the door.
Close DOOR tightly.
Press RUN/PAUSE again.
Listen for 3 beeps and look for RUN on the display; the pump is running.
Message on display screen: SHUT DOOR
Solution: Press RUN/PAUSE
Close pump door tightly; press RUN/PAUSE.
Listen for 3 beeps and look for RUN on the display; the pump is running.