This chapter describes routine maintenance procedures for your RC-5C PLUS Centrifuge. It also includes
cleaning and decontamination recommendations and the Service Decontamination Policy. As the user, it
is your responsibility to make certain these procedures and recommendations are followed. In addition to
these routine procedures, to keep your centrifuge in good working condition and ensure accurate test
results, we recommend that you have the centrifuge serviced and have the speed control, timer, tempera-
ture control, and rotor imbalance checked annually by a Thermo Field Service Engineer or other qualified
service personnel. If further service is needed, contact your Thermo Field Service Engineer.
W A R N I N G
!
There are no user-service-
able items inside the centrifuge. Be-
cause of the high voltages in the cen-
trifuge, anyone who is not properly
trained in electronics must not test or
repair the electrical circuits.

Maintenance

To gain the most in safety, performance, and useful life from your
centrifuge, it is necessary to properly maintain it. Proper mainte-
nance is a simple and inexpensive way to help prevent major fail-
ures. In addition to the maintenance described below, a Thermo
service contract, which includes a yearly Preventive Maintenance
visit, is recommended to keep your centrifuge in good condition.
Table 5-1 suggests a maintenance schedule.
Table 5-1. Maintenance Schedule
Maintenance
Clean the rotor chamber
Wash the tapered spindle
Clean the cabinet panels
Have condenser fins cleaned
Check LEDs and indicators
Have ground continuity tested
Clean the air inlet on the drive
box (wipe with cloth)
Care and Maintenance
Frequency
Daily or immediately after a spill
Once a week
Once a month
Once a year or whenever centri-
fuge is serviced; more often if dust
accumulates significantly
Once a year or whenever the dis-
played numbers seem erroneous
Once a year or whenever centri-
fuge is serviced
Once a month
5-1