Install Option SS4 (Instruments with Front Panel RF Output)
Install Option SS4 (Instruments with Front Panel RF Output)
Remove Existing A19 40 GHz Doubler
1. Refer to
Figure
3. Remove cables W23 and W25. Also disconnect the cable at the location
indicated.
Figure 3
A19 40 GHz Doubler Cables and Screws
2. Remove the four screws (1) (0515-0372) that attach the A19 40 GHz Doubler to the deck. The
40 GHz Doubler can now be lifted from the deck.
Install A19 40 GHz Doubler
1. Reverse the order of the removal procedure.
2. Recommended torque for all screws is 9 in–lbs.
3. Recommended torque for all cable connectors is 9 in–lbs.
8
Installation Note N5191-90021