Measuring AC Current (MP730007 only)

Insert the black test lead into COM terminal. If the current under test ≤ 200mA, insert
the red test lead into VΩmA terminal. If the current is higher, insert the red test lead into
the 10A terminal.
Turn the range selector to AC/DC range, and connect the test probes with the
circuit being measured in series.
Notes:
If the current to be measured is unknown, choose the maximum measuring range
(10A) and reduce it step by step until a satisfactory reading is obtained.
Disconnect from the circuit under test when changing the range.
If the display only shows "OL", it indicates the measured current is out of range. In
this case, turn the range selector to a higher range.

Measuring Resistance

Insert the test leads into the input terminals (red to VΩmA and black to COM).
Turn the range selector to Ω range, and connect the test probes to the resistor
being measured.
Notes:
If the measured resistance is out of the range selected, "OL" will be displayed and
you should select a higher range. It normally takes several seconds to display a
stable reading in a high-resistance (>1MΩ) measurement.
When there is no input, such as in open-circuit condition, the instrument will
display "OL".
When making an in circuit measurement, disconnect all power supplies and
discharge all the capacitors.
When measuring low resistance, the test leads will produce 0.1Ω-0.2Ω
measurement error. To obtain accurate measurement, short the test leads and use
REL function.
If the resistance when shorted is more than 0.5Ω, please check if test leads are
loosened or damaged.
Do not input over 60V DC or 30V AC or it will pose shock hazard.

Testing Continuity

Insert the test leads into the input terminals (red to VΩmA and black to COM, the
polarity of the red test lead is "+").
Turn the range selector to continuity range.
Connect the test probes with the load in parallel. The reading is displayed.
If the measured resistance
If the measured resistance
sound.

Diode Testing

Switch the
range selector
Insert the red test lead to
Red test lead to positive pole, black to negative pole. Reading is displayed.
"OL" symbol appears when the diode is open or polarity is reversed. For silicon PN
junction, normal value: 500 - 800mV (0.5 - 0.8V).
Notes:
The voltage for testing diode is about 4.0V/1.5mA.
>51Ω
indicates an open-circuit.
<100Ω
the circuit is in good conduction and buzzer will
to diode position
VΩmA
terminal, black to COM terminal.