multicomp pro MP730989 User Manual

Smart digital multimeter

Advertisement

Quick Links

MP730989
Smart Digital Multimeter User Manual
Overview
MP730989 is a 9999-count true RMS digital multimeter, with full push-button design and laser-engraving
light indication, ii is clearly visible and easy to be operated in dark environments. UT60S can be used to
measure large capacitance up to 99.99mF and automatically identify current input terminal. The meter has
multiple features such as overvoltage alarm indication, overcurrent alarm indication, false detection
protection for high voltages, etc.
Features
• Full push-button design and laser-engraving light indication.
• Automatically identify current input terminal.
• The power frequency value shows on the slave display when measuring AC voltage.
• 9999-count display, true RMS measurement, and fast ADC (3 times/s).
• Full-featured false detection protection for up to 1000V surge, and overvoltage/overcurrent alarm.
• Extended measuring range, the � 1 00mF response time for stabilizing reading is within 10 seconds compared
with similar products.
• NCV function, audible and visual alarm.
• Measurement of fiame sensor of heating device can be performed at uA position
• Low power consumption (general: 7mA; sleep state: 1 OuA) to effectively extend the battery life to 300 hours.
Accessories
Open the package box and take out the meter . Please double check whether the following items are missing
or damaged.
1. User manual --------1 pc
2. Test leads ----------1 pair
3. Temperature probe------ 1 pc
If any of the above is missing or damaged, please contact your supplier immediately.
&
Read the "Safety Instruction" carefully before use.
Safety Instruction
1. Safety Standards
1) The meter is designed according to EN 61010-1:2010; EN 61010-2-030:2010; EN 61010-2-033 :2012 and
EN 61326-1:2013; EN 61326-2-2:2013.
2) The meter conforms to, double insulation, CA T II 1000V/CAT Ill 600V overvoltage standard, and pollution
degree 2.
2. Safety Information
1) If use the meter without following up the operating instructions, the protection provided by the meter may be
impaired or lost.
2) Do not use the meter if the rear cover is not completely covered up, or ii may pose a shock hazard and the
meter is designed for indoor use.
3) Check and make sure the insulation of the meter and test leads is in good condition without any damage
before use. If the insulation of the meter casing is found to be significantly damaged, or if the meter is
considered to be malfunctioning, please do not continue to use the meter.
4) Keep fingers behind the finger guards of the test leads when using the meter.
5) Do not apply more than 1 OOOV between any terminal and earth ground to prevent electric shock and damage
to the meter.
6) Use caution when working with voltages above AC 30Vrms or DC 60V. Such voltages pose a shock hazard.
7) The measured signal is not allowed to exceed the specified limit to prevent electric shock and damage to
the meter.
8) Place the function dial in the correct position before measurement.
9) Never turn the function dial during measurement to avoid damage to the meter.
1 O) Do not change the internal circuit of the meter to avoid damage to the meter or user.
11) Damaged fuses must be replaced with fast-acting ones of same specifications.
12) When· 11".D" is displayed, please replace the batteries in time to ensure measurement accuracy.
13) Do not use or store the meter in high temperature, high humidityfiammable, explosive, or strong magnetic
field environments.
14) Clean the meter casing with a damp cloth and mild detergent. Do not use abrasives or solvents.
15) Use of test probe.
TESTING IN CAT Ill/IV MEASUREMENT LOCATIONS
Ensure the test lead shield pressed firmly in place. Failure to use the CAT Ill/IV shield increase arc-fiash risk.
TESTING IN CAT II MEASUREMENT LOCATIONS
CAT Ill shields may be removed for CAT II locations. This will allow testing on recessed conductors such as
wall outlet. Take care not to lose the shields.
��
1§1
C€
Electrical Symbols
CE
&.
Caution, possibility of electric shock
,,...,
Alternating current
-
---
.&.
Direct Current
Equipment protected throughout by DOUBLE INSULATION or REINFORCED INSULATION
[g]
- � -
· �
Conform to UL STD 61010-1, 61010-2-030, 61010-2-033, certified by CSA STD C22.2 No.
61010-1, 61010-2-030, 61010-2-033
It is applicable to test and measuring circuits connected to the distribution part of the building's
CATIII
low-voltage MAINS installation.
It is applicable to test and measuring circuits connected directly to utilization points (socket
outlets and similar points) of the low-voltage MAINS installation.
CATII
General Specifications
1. The maximum voltage between input terminal and earth ground is 1 OOOVrms.
2. 10A terminal protection:10A H 1000V quick-acting fuse, <!>6•32mm, Fuse breaking rating: 10KA
3. 9999-count display, show "OL" when overrange, update 3 times per second.
4. Range: Auto
5. Backlighl: manually turn on and auto turn off after 30 seconds.
6. Polarity: Display symbol "-" for negative polarity input.
7. Data hold: "HOLD" display on top right of LCD.
8. Low battery indication: "P!.!I" display on bottom left of LCD.
9. Battery: AAA battery 1.5V•3
10. Working temperature: o·c~40'C (32
F-104
F)
°
°
Storage temperature: -1o·c~so·c (14"F~122'F)
Relative humidity: o·c~30'c .,;rs%, 30'C~40'c .,;so%
Working altitude: o-2ooom
11. Dimension: 169*81*46mm
12. Weight: about 290.2g (including batteries)
External structure (Figure 1)
1. LCD display
2. Audible and visual indicator
3. Function button
4. Input terminal
5. Hanging hook
6. Test probe holder
7. Battery cover
8. kickstand
Function buttons
*. S : 1. Short press to enter manual range switch mode for voltage, current and resistance.
2. Long press for �2s to return to auto measurement mode.
3. Press for ?2s when turning on the meter to disable auto power off function. (Full display: POFF)
'. !S:
1. The sub window under ACVIACA function displays frequency measurement.
2. Short press to display frequency in main window and display duty cycle % in sub window.
'. EJ:
1. Short press HOLD to hold current test data, "HOLD" is shown on LCD.
2. Long press HOLD for about ?2s to enter REL mode, LCD displays"�".
•. l.:'i' :
Long press (?2s) to turn on the meter, long press again to turn it off.
•. ll!! :
Long press to switch input signal of AC/DC voltage and NCV.
'. � : Long press to switch signal of AC/DC voltage and temperature.
•. !iIJ :
Switch input signal of resistance/continuity/capacitance/diode and retain current function.
•. (iil : Automatically identify current after the test probe is connected, in current mode, short press to switch
AC/DC current measurement functions.
Operating Instructions
1. AC/DC Voltage Measurement (Figure 2)
1) Short press " � " for AC/DC voltage test.
2) Short press"� " for AC/DC mV voltage test
3) Insert the red test lead into the "VO" terminal, black test lead
load).
4) Read the test result from LCD.
&warning
• Do not input a voltage over 1 OOOV, or ii may damage the meter
and hurt the user.
• The input impedance of the meter is 10MO. This load effect may
impedance of the circuit is ,; 1 OkO, the error can be ignored
(s0.1%).
• Be cautious to avoid electric shock when measuring high voltages.
• Before each use, verify meter operation by measuring a known
voltage.
2. Resistance Measurement (Figure 3)
1) Short press "
2) Insert the red test lead into the "VO" terminal, black test lead into the "COM" terminal, and make the probes
3) Read the test result from LCD.
&warning
•. Before measuring resistance, switch off the power supply of the
•. If the resistance is not less than 0.50 when the test leads are
•. If the measured resistor is open or the resistance exceeds the
•. When measuring low resistance, the test leads will produce
0.10-0.20 measurement error. To obtain the final accurate value,
*. When measuring high resistance, it is normal to take a few
*. Do not input voltages over 60 VDC or 30 VAC.
3. Continuity Test (Figure 3)
1) Short press "
2) Insert the red test lead into the "VO" terminal, black test lead into the "COM" terminal, and make the probes
3) When measured resistance.,; 100, the circuit is in good conduction status and the buzzer beeps continuously
& warning
*.Switch off the power supply of the circuit and discharge all capacitors before test.
4. Diode Test (Figure 3)
1) Short press "
2) Insert the red test lead into the ''VO+!-" terminal, black test lead into the "COM" terminal, and make the probes
3) If the diode is open or its polarity is reversed, the LCD will display "OL". For silicon PN junction, the normal
& warning:
•. Switch off the power supply of the circuit and discharge all capacitors before testing the PN junction.
•. The test voltage is about 4.0V/1.5mA.
5. Capacitance Measurement (Figure 3)
1) Short press
2) Insert the red test lead into the "VO" terminal, black test lead into the "COM" terminal, and make the probes
3) When there is no input, the meter displays a fixed value (intrinsic capacitance). For small capacitance
Comply with European Union directives
&warning
• If the measured capacitor is shorted or the capacitance exceeds the maximum range, the LCD will display "OL".
Earth (ground) TERMINAL
• When measuring high capacitance, ii is normal to take a few seconds to stabilize the reading .
• Before measuring, discharge all capacitors (especially high-voltage capacitors) to avoid damage to the meter
Warning
and user.
6. AC/DC Current Measurement (Figure 4a,4b)
1) Insert the red test lead into the "µA" or "mA/A" terminal, black test lead into the "COM" terminal.
2) Short press " fj;l " to switch between AC and DC current.
3) Connect the test leads to the power supply or circuit under test in series.
4) Read the test result from LCD.
&warning
• Switch off the power supply of the circuit, make sure the input terminals and dial position are correct, and then
connect the meter to the circuit in series.
• If the range of the measured current is unknown, select the maximum range and then accordingly reduce.
• If the "mNA" terminal is overloaded, the built-in fuse will be blown and must be replaced. Self-recovery
protection circuit is built for the overload of "uA" terminal.
• Do not connect the test leads to any circuit in parallel during current measurement to avoid damage to the
meter and user.
• "µA" terminal can be used to heating system flame rectifier circuit detection (Figure 4b)
• When test leads did not inserted into terminal and press " EiJ" " current button LCD would display "LEAd" to
7. Temperature Measurement (Figure 5)
1) Long press " -
2) Insert the K-type thermocouple into the ''VO" and "COM" terminals, and fix the temperature sensing end of
,&
The LCD displays "OL" when the meter is turned on. Only K-type thermocouple is applicable, and the measured
temperature should be less than 250
into the "COM" terminal, and make the probes in contact with
both ends of the measured voltage (parallel connection to the
cause measurement errors in high-impedance circuits. If the
GJ "
for resistance test.
in contact with both ends of the measured resistance (parallel connection to the resistance).
circuit and discharge all capacitors before measuring resistance.
shorted, please check if the test leads are loose or abnormal.
maximum range, the LCD will display "OL".
the resistance of shorted test leads should be subtracted from
the measured resistance value.
seconds to stabilize the reading.
E.J"
for continuity test.
in contact with the two test points.
along with a green LED indication. When measured resistanc.,,.500, the circuit is broken, buzzer makes no
sound along with a yellow LED indication. When measured resistance ?1000, LED shows red indication.
9"
for diode test.
in contact with the two endpoints of the PN junction.
value is generally about 500mV-800mV (0.5V-0.8V). Buzzer beeps instantly when readout displayed, test
leads short-circuit buzzer beeps continuously.
"mil"
for capacitance test.
in contact with the two endpoints of the capacitance.
measurement, this fixed value must be subtracted from the measured value to ensure measurement accuracy.
So, please use the relative value measurement (REL) mode to automatically subtract the fixed value.
prompt to insert test leads
@) )(
Q Load
Figure4a
� " for temperature test.
the thermocouple on the object under test, read the temperature from LCD after the value stabilizes.
(:( l
////
K
Figure 5
Warning
C/482
F (
F = ·c • 1.8
°
°
°
+
Figure 2
Figure 3
Control
Module
Flame
sensor
probe
Figure 4b
32).

Advertisement

loading
Need help?

Need help?

Do you have a question about the MP730989 and is the answer not in the manual?

Questions and answers

Summary of Contents for multicomp pro MP730989

  • Page 1 "COM" terminal, and make the probes in contact with MP730989 is a 9999-count true RMS digital multimeter, with full push-button design and laser-engraving light indication, ii is clearly visible and easy to be operated in dark environments. UT60S can be used to both ends of the measured voltage (parallel connection to the measure large capacitance up to 99.99mF and automatically identify current input terminal.
  • Page 2 8. Frequency Measurement (Figure 6) · e " 4. Continuity and Diode 1) When measuring AC voltage/current, press" to enter frequency or duty cycle measurement mode. Range Resolution Accuracy 2) Insert the red test lead into the ''VOHz" terminal, black test lead into the "COM" terminal, and connect the Broken circuit: Resistance 2'500, no beep, light up yellow.