Skylife Operating Instructions; Setting Up The System - NeoLight SKYLIFE User Manual

Phototherapy system
Hide thumbs Also See for SKYLIFE:
Table of Contents

Advertisement

5. SKYLIFE OPERATING INSTRUCTIONS

(Steps 1 - 6 are applicable in both home and hospital settings.)

5.1 SETTING UP THE SYSTEM

1
Place the Skylife device in a location with walls
(e.g. bassinet, warmer, crib, or incubator).
Caution: Do not allow items such as blankets to cover
the outside edges of the device.
2
Visually inspect Light Bed and GelMat
for damage or wear. If GelMat becomes
damaged, torn, or develops a yellowish
discoloration, refer to section 6.2, steps
#1 - 4 for replacement information. A home
user should contact the system provider for
replacement.
3
Place the Cloud Swaddle on the device
over the GelMat. Secure the Cloud Swaddle
to the device using the corner pockets like a
fitted bed sheet.
4
Attach the Controller and Power Supply
to the device and place the Controller
outside the enclosing walls of the treatment
area. Ensure that the cord does not pose a
hazard to the baby. Plug the Power Supply
into the wall outlet.
500007, Rev P, Pg 6