High Temperature Alarm Condition:
System Failure Condition:
Accidental Power Disconnection:
POWER INTERRUPTED
RESUMING TREATMENT IN 09 SEC
Settings:
LED LIFETIME
VERSION #
** LED lifetime refers to the number of hours the LED modules have been used to date.
If the system temperature increases beyond a certain threshold, the High Temperature
Alarm will sound and LED indicators on the controller will flash. Press any button except
power button to mute the alarm. Check if the air vents on the underside of the device
are blocked, remove any items that may be blocking air vents. Resume treatment.
If high temperature condition persists, power OFF the device and remove the baby from
the Skylife device. Contact treating physician, technician, or manufacturer.
If any part of the device fails to perform as expected, a "System Failure' alarm will sound,
LED indicators on controller will flash and the device will shut OFF automatically. Restart
the device to see if it works. If the device shows the same error message or fails to power
on, contact treating physician, hospital technician, or manufacturer.
If the power cable is accidentally unplugged from the wall socket, plug in the power
cable back into the socket. The system will display a count down prior to automatically
continuing treatment. To pause treatment, press the power button . To discontinue
treatment, press and hold the power button.
Caution:
The system will automatically restart treatment within 10 seconds of power
application.
Press Settings button to view the system settings. The Settings will display following
information:
LED Lifetime**
Version Number
Only a certified technician may change these settings