Parameters Setting; Measurement; Standard Measurement; Minimum Value Measurement - DGC DC-2000C Series Instruction Manual

Ultrasonic thickness gauge
Table of Contents

Advertisement

DC-2000C SERIES ULTRASONIC THICKNESS GAUGE
【8.4】Parameters setting
〖8.4.1〗 Measurement
There are six measuring modes provided. Users can select different measuring
modes according to their requirements and measuring environments.
- Press
- Press
- Press
- Press

8.4.1.1 Standard measurement:

Display the current value, satisfied with the normal measuring needs.

8.4.1.2 Minimum value measurement:

Among one measurement, display the minimum value of the current measured
point. It is suitable for testing the curvature surface or needs to get the
minimum value which is widely used in the thickness measurement of
pipeline.
Notice: It is not recommended to use this function when measuring cast
iron or alloy materials
8.4.1.3 Difference measurement: (DC-2020C)
Display the accurate differential value between the measured value and
reference value set by the users, suitable for quality check to identifying the
qualified products whose thickness is in the admissive error or not.
8.4.1.4 Average mode: (DC-2020C)
Provides the average value of 2 to 9 measured points and display it, suitable
for testing the flat surface.
8.4.1.5 Limitation setting: (DC-2020C)
Set the upper and lower limit, when the measured thickness exceeds the preset
1. Measurment
into
or
to select desired measurment mode.
to confirm,
to Esc. Menu and into measurment state.
8

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dc-2020c

Table of Contents